Commands and Tips for Third-Party Devices
Gilson 321 and 322 Pumps
In addition to the standard pump commands (see Commands for Controlling Dionex Devices Dionex Pumps), the following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
Property |
(Min/Max) Values |
Description |
ModelNo |
N/A |
Indicates the model number of the pump (read-only) |
Firmware Version |
x.xx |
Indicates the pump's firmware version (read-only) |
ChartOut (321 only) |
Possible values: Pressure, Flow, Percentage %A, %B |
Determines the property to be monitored via the digital output (pin no. 9). |
|
|
|
Pump Properties |
||
|
|
|
%A_RefillTime, %B_RefillTime |
125 –1000 ms |
Determines how long the piston takes to return to its initial state. Enter a higher value in case of cavitation. |
%A_LiquidCompr %B_LiquidCompr |
0 – 2000 1/Mbar |
Determines the solvent compressibility. |
|
|
|
Prime Commands |
||
|
|
|
StartPriming |
PumpHead parameter (possible values: RightHead, LeftHead) |
Starts a priming sequence for the specified pump head. Note: After performing the StartPriming command, no other command or property is allowed until the StopPriming command was performed. |
StopPriming |
- |
Finishes a running priming sequence. |
|
|
|
Maintenance Counters |
||
|
|
|
Right_PistonSeal, Left_PistonSeal |
0 - 65534 h |
Maintenance counter for the right or left piston seal (read-only) |
Right_Piston, Left_Piston |
0 - 65534 h |
Maintenance counter for the right or left piston (read-only) |
Right_OutletValve, Left_OutletValve |
0 - 65534 h |
Maintenance counter for the right or left outlet valve (read-only) |
Right_InletValve, Left_InletValve |
0 - 65534 h |
Maintenance counter for the right or left inlet valve (read-only) |
Right_Actuator, Left_Actuator |
0 - 65534 h |
Maintenance counter for the right or left actuator (read-only) |
Further Information
Tip:
For more information, for example, about the empirical compressibility correction values for the Gilson 321 and 322 pumps, refer to the pump's User's Guide.
For information about how to install these pumps, refer to Hardware Installation Gilson 321 and 322 Pumps: Installation in the Administrator Help section.