Dionex Detectors
Dionex/LC Packings UltiMate Capillary/Nano HPLC UV Detectors

In addition to the standard commands available for all detectors (see  Dionex Detectors), the UltiMate detector provides the following control commands (please note that the display Filter level determines which commands and properties are displayed):

 

Command/Property

Description

Adjust_Calibration

Performs wavelength calibration by using an internal test.

 Tip:  

The command stops the data acquisition.

Data_Collection_Rate

Enter the rate at which Chromeleon collects digital data from the detector.

DemoFileName

Name of the file that is read in the Demo Mode.

LampType

Indicates the lamp type: Deuterium or Tungsten (read-only).

TimeConstant

Time constant. This is the time the detector requires for rising from 0% to 63% in response to a signal step. The constant is used for noise suppression (values: 0.0 s to 4.55 s; depending on the detector type only a predefined scale of values may be available). The time constant is directly associated with the Rise_Time. The following formula applies: 2.2 x TimeConstant = Rise Time.

 

 Tip:  

Most noise is suppressed if the time constant is 4.55 s. However, for the acquisition of narrow peaks a lower response will be more appropriate.

 

 Tip:

If you perform the Data Collection Rate command, the default Step value is set to the reciprocal value. That is why you have to perform the Step command after the Data Collection Rate command if you need a different step.

The UV_Diagnostic section provides information about the detector operation and service. The entries are read-only:

 

Command/Property

Description

LampAge

Reports the total operating hours of the lamp.

LampIgnitions

Reports the number of lamp starts.

LampNumber

Indicates the lamp number. (The value is incremented when a new lamp is installed.)

LampOnTime

Reports how many hours the lamp has been burning since it was turned on.

LastServiceDate

Reports when the detector was last serviced. (The Dionex Service Engineer enters the date.)

LastServiceCode

Reports the service code, providing information about the service. (The Dionex Service Engineer sets the code.)

ManufacturingDate

Reports the detector manufacturing date.

MotorWorkTime

Reports the total operating hours of the motor.

 

For information about how to install the UltiMate UV detector, refer to Hardware Installation  UltiMate Capillary/Nano HPLC System: Installation.