Dionex Detectors
Dionex CD20/CD25/CD25A, ED40/ED50/ED50A, and IC20/IC25/IC25A Electrochemical Detectors

In addition to the standard commands available for all detectors (see  Dionex Detectors), Dionex electrochemical detectors provide special commands (see the table below; please note that the display Filter level determines which commands are displayed). The Data_Collection_Rate command and the analog output commands are available for all Dionex electrochemical detectors in any operating mode. The Mode command is available only for the ED40/ED50/ED50A detectors. Other special commands depend on the selected operating mode.

 

Command

Description

Data_Collection_Rate

Sets the rate at which Chromeleon collects digital data points from the detector.

Mode

(Available only for the ED40/ED50/ED50A) Select the operating mode: conductivity, integrated amperometry, or DC amperometry.

 

Analog Output Commands (All Modes)

 

Command

Description

Full_Scale

Sets the full-scale analog out voltage.

Mark

Sends a positive pulse to the detector’s analog output as an event marker.

Offset_Level

Sets the offset for the analog output signal.

Polarity

Switches the analog output polarity (+/-).

Recorder_Range

Sets the range of a full-scale recorder response.

Rise_Time

Sets how quickly the detector responds to a change in signal.

 

Special Commands (Conductivity Mode)

The following commands are available for conductivity detectors (CD20/CD25/CD25A, ED40/ED50/ED50A, or IC20/IC25/IC25A):

 

Command

Description

DS3_Temperature

Sets the DS3 temperature.

SRS_Current

For CD20/CD25, ED40/ED50, IC20/IC25 only: Sets the current supplied to the SRS.

Suppressor_Current

For CD25A, ED50A, and IC25A only: Sets the current supplied to the suppressor.

Suppressor_Type

For CD25A, ED50A, and IC25A only: Sets the installed Suppressor.

Temperature_Compensation

Sets the temperature compensation factor, which is used to stabilize conductivity readings.

 

Special Commands (DC Amperometry Mode)

The following commands are available for the ED40/ED50/ED50A detectors in DC amperometry mode:

 

Command

Description

Cell

Turns the cell on and off.

DC_Voltage

Sets the fixed potential applied to the working electrode, in the range of -2.04V to +2.04V.

Electrode

Select the type of electrode installed.

 

Special Commands (Integrated Amperometry Mode)

The following commands are available for the ED40/ED50/ED50A detectors in integrated amperometry mode:

 

Command

Description

Cell

Turns the cell on and off.

Electrode

Select the reference electrode type.

Waveform

Define a waveform (a plot of potential vs. time).

 

Calibration Commands (For CD25A, ED50A, and IC25A Only)

 

Command

Description

LeakDet

Performs the leak detector calibration.

CondCell

Performs the conductivity cell calibration

DownloadCalibration

Downloads the selected set of calibration values to the detector.

DownloadCalibration Parameter

Selects the set of calibration values (Current, Previous, or Factory) to be downloaded.

 

Diagnostic Commands (For CD25A, ED50A, and IC25A Only)

 

Command

Description

LeakDetTest

Performs a test of the leak detector.

LeakDetTestDate

Reports the date of the leak detector test.

LeakDetTestResult

Reports the result (pass or fail) of the leak detector test.