Dionex Autosamplers
Dionex/LC Packings FAMOS Autosampler: Sample Preparation
The Dionex/LC Packings FAMOS Autosampler provides many sample preparation commands that are used to further define the individual sample preparation steps. The sample preparation commands are accessible from the Advanced level only. Use the Program Wizard to create a user-defined sample preparation program. This program is started by the Inject command.
Chromeleon supports the following sample preparation commands (please note that the display Filter level determines which commands and properties are displayed):
Command/Property |
Description |
CodeOutput |
Communicates a code number (between 0 and 15) to the P2 output (VialNo&Marker (TTL)). For more information, refer to the FAMOS User's Manual. |
Compressor |
Serves to build up pressure (On) or relieve pressure (Off) |
Dispenses a specified volume. The following options are available: To: Sample position to which to dispense the volume. Volume: Volume to dispense (0.0 to 25.0 µl). SyringeSpeed: Speed used to dispense the volume (Either select an option (Low, Normal, or High) from the drop-down list or enter the any number between 1 and 9). SampleHeight: Height at which the sample is dispensed (0 to 40 mm). |
|
Draws a specified volume. The following options are available: From: Sample position from which to draw the volume. Volume: Volume to dispense (0.0 to 25.0 µl). SyringeSpeed: Speed used to draw the volume (Either select an option (Low, Normal, or High) from the drop-down list or enter the any number between 1 and 9). SampleHeight: Height at which the sample is drawn (0 to 40 mm). |
|
InjectMarker |
Required to start injection. |
InjectValve |
Injection valve. The valve can be either in Inject or in Load position. |
LabeledVialMarker |
Creates another marker pulse for different vials, e.g., standard samples. |
MixWait |
Specify the time the system waits between steps |
MixNeedleWash |
Volume used for washing the needle (10 to 9.999µl). |
MoveSyringe |
Determines the volume and the speed for loading and dispensing the syringe. For this command, the Load and Unload parameters are available. Tip: Do not use both parameters in the same line of the PGM File because in this case, the PGM File does not pass the Ready Check. |
MoveSyringeHome |
Returns the syringe to its home position. |
Output |
Sets the corresponding output to the selected state. |
SSV |
Solvent Selection Valve. Specify the Position. The following options are available: Solvent A-D. |
SyringeValve |
Specify the Position. The following options are available: Needle, Wash, or Waste. |
Valve_A |
Specify the Position. The following options are available: |
Valve_B |
Specify the Position. The following options are available: |
VialMarker |
Generates a marker pulse for sample vials, e.g., unknown samples. |
WaitInput |
Defines the incoming signal for which to wait. |
Starts a wash cycle. |
For a basic description of Dionex autosamplers, including the standard commands, refer to Dionex Autosamplers.
For information about the general commands for the FAMOS autosampler, refer to Dionex/LC Packings FAMOS Autosampler. (The general FAMOS autosampler commands partly overlap the sample preparation commands listed above.)
For information about how to use the commands in practical operation, refer to Practical Tips for Device Control Examples for User-Defined Programs for the FAMOS Autosampler (Dionex/LC Packings).