Dionex Sample and Fraction Manager
Sample and Fraction Manager Commands
Apart from General Device Commands, the Sample and Fraction Manager supports the following commands and properties (please note that display Filter level determines which commands and properties are displayed):
Status properties
Property |
Value Range |
Filter |
Description |
Busy |
Busy/Idle |
Expert |
Indicates whether the device driver has not yet sent all commands or whether execution of a command is yet complete |
CPUSerialNo |
- |
Expert |
Indicates the CPU serial number of the connected sampler |
DeviceState |
Ready/Busy/ |
Normal |
Indicates the current device status (read-only). |
StatusDescription |
- |
Normal |
Indicates the current action of the sampler. If no action is performed, Idle is displayed |
Syringe |
- |
Normal |
Indicates the installed syringe type (read-only). |
Inject properties
Command |
Value Range |
Description |
AirVolume |
Depends on the volume of the installed syringe; indicated in µl. |
Specifies the air volume that is drawn for the Inject and/or WashWithSample commands (in LC_Inject and LC_Overlap mode). |
FillSpeed |
Depends on the minimum and maximum speed of the installed syringe; indicated in µl/s. |
Fill speed for the preparing fill strokes for the Inject and/or WashWithSample commands. |
FillStrokes |
0-99 or Default* |
Number of preparing fill strokes for the Inject and/or WashWithSample commands (in LC_Inject and LC_Overlap mode). |
FillVolume |
Depends on the syringe volume; indicated in µl. |
Fill volume for the single piston strokes (for the WashWithSample command). |
InjectMode |
LC_Inject |
Injection mode. In sample programs, this value must be set prior to the Inject command. |
Injector |
List of injection objects for the respective arm of the installed Sample and Fraction Manager |
Injector to be used for the Inject and/or WashWithSample commands (in LC_Inject and LC_Overlap mode). The setting is ignored in Custom injection mode. |
InjectSpeed |
Depends on the minimum and maximum speed of the installed syringe; |
Syringe speed in µl/s for injections to the sample loop |
NoOfPlungerStrokes |
|
Indicates the number of strokes since the last reset (Value) and maximum allowed number (Max.) as reported by the device. (The property is read-only.) |
NoOfInjects |
|
Indicates the number of injections since the last reset (Value) and maximum allowed number (Max.) as reported by the device. (The property is read-only.) |
Penetration |
0 - maximum value of all installed tray types |
Defines how deep the needle descends into the vial (parameter for Inject, PrepareThisSample, PrepareNextSample, WashWithSolvent). |
Position |
1 to maximum number of tubes allowed for the trays used in the current configuration. |
Defines the absolute position via an index entry, which can be defined as <TrayName> <Tube>. Note: When you select a different position, the Tube, Tray, and TrayName properties are updated automatically. |
PostInjectDelay |
0 - 99000 ms or Default* |
Time between the injection and the removing of the needle from the injector (in LC_Inject and LC_Overlap mode). |
PreInjectDelay |
0 - 99000 ms or Default* |
Time between inserting the needle in the injector and injecting the liquid into the sample loop (in LC_Inject mode). |
PullupDelay |
0 - 10000 ms or Default* |
Indicates the delay time between drawing the liquid into the syringe and dispensing the volume that is not required (for the WashWithSample command). |
Tray |
|
Tray used (read-only; determined via the Position property).
|
TrayName |
String |
Indicates the name of the current tray Note: If you select a different tray, the Tray property is updated automatically. |
Tube |
|
Number of the current tube in the tray (read-only; determined via the Position property) |
* Note:
For the Default parameter, the default value set on the respective device is used.
Command for the analytical SFM
Command |
Description |
CloseStack |
Closes all stacks that might be open. This might be useful when fraction collection is finished. |
Commands for the LC_Overlap mode
Command |
Description |
PrepareNextSample |
The autosampler fills the sample loop with the next sample, observing the parameters defined in the PGM file for that sample (Volume, Position, Tray, AirVolume, FillSpeed, FillStrokes, and PullupDelay). The syringe then returns to the Park position. This command is ignored for InjectModes other than LC_Overlap. This command is available only after the Inject command in the PGM File. |
PrepareThisSample |
The autosampler fills the sample loop with the current sample, observing the parameters defined in the PGM File for the current sample (also, see PrepareNextSample). This command is available only before the Inject command in the PGM File. |
For more information, refer to: