Installing and Controlling Third-Party Devices
Soma 3710 UV Detector: Installation
Hardware Installation
Software Installation
Install the Device Driver in the Chromeleon Server Configuration. (For more information about the program, refer to Software Installation and Communication The Server Configuration Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the detector to the timebase. Select Add Device on the Edit or context menu, and then select Soma from the Manufacturers list box and Soma_UV-Vis_Detector_S-3710_178 (up to firmware version 1.78) or Soma_UV-Vis_Detector_S-3710_New (from firmware version 1.79) from the Devices list box.
The following settings are required:
On the Soma_UV-VIS_Detector_S-3710_178 tab page (Soma_UV-VIS_Detector_S-3710_New tab page, respectively), disable the Demo Mode.
On the UV tab page, enter a name for the UV detector. In addition, select a free COM port to which the detector is connected and define the channel for data acquisition.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the installed Soma 3710 UV Detector, connect the UV-Fluorescence.pan Control Panel or any especially developed control panel in the Chromeleon client to the corresponding timebase.
Application
Nothing particular known.
Further Information
Tip:
For more installation details, refer to the manual for the instrument.
For an overview of the detector, refer to Soma 3710 UV Detector: Overview.