Installing and Controlling Third-Party Devices
Shimadzu LC-10A/LC-20A Prominence HPLC Systems: Pumps

Chromeleon supports the following pump combinations (the bold entries in the table indicate the commonly used pump mode). Chromeleon also supports up to 3 FCVs. Any installed FCV can be used as a solvent selection valve, with the exception of FCV.A in LP.GE mode.

 

Installed Pumps

LPG Application

HPG Application

Pump Mode (Shimadzu terminology)

1 pump
no FCV.A

(Only %A possible)

-

ISO

1 pump, FCV.A²

Only LPG with up to 4 channels possible.

-

ISO or

LP.GE

2 pumps
no FCV.A

(Only %A possible)

HPG with up to 2 channels possible.

ISO + Pump.B or

B.GE

2 pumps +
FCV.A

LPG with up to 4 channels possible.

HPG with up to 2 channels possible.

ISO + Pump.B or

LP.GE + Pump.B or

B.GE

3 pumps
no FCV.A

(Only %A possible)

HPG with up to 3 channels possible.

ISO + Pump.B + Pump.C or

B.GE + Pump.C or

T.GE

3 pumps +
FCV.A

LPG with up to 4 channels possible.

HPG with up to 3 channels possible.

ISO + Pump.B + Pump.C or

LP.GE + Pump.B + Pump.C or

B.GE + Pump.C or

T.GE

² For low-pressure gradient applications, a flow channel valve (FCV) must be connected to pump A.

Abbreviations used in the table:

FCV = Flow Channel Valve

HPG = High-Pressure Gradient

LPG = Low-Pressure Gradient

Make the necessary entries on the Pumps tab page in the  Server Configuration program.

  Tip:

If you have selected an option for which one or several modules are not available, the mode is automatically set to Isocratic and an error message is displayed:

 

For information about how to install the HPLC system, refer to  Shimadzu LC-10A/LC-20A HPLC System: Installation.

 

For application information, refer to:

Shimadzu LC-10A and LC-2010 HPLC Systems: Pump

Shimadzu LC-10A/LC-20A/LC-30A HPLC Pumps: Tips