Installing and Controlling Third-Party Devices
Perkin Elmer TurboMatrix Headspace Sampler: Installation
Device Type: |
GC Headspace Autosampler |
Device Driver: |
Perkin Elmer: TurboMatrix Headspace Sampler |
Hardware Options |
|
Restrictions: |
The following options and/or versions are not supported:
|
What is required?
Hardware Installation
Device Connection: |
Connect the sampler to the PC, using a null modem cable with 9-pin connector (Dionex part no. 1310.2263, see Null Modem Cable in the Appendix of the Hardware Installation section). Perkin Elmer provides different cables for connecting the sampler to a GC: Autosystem (XL) and Clarus 500: Perkin Elmer part no. L427-0032 Agilent 5890/6890: Perkin Elmer part no. N610-0410 Other GCs: Perkin Elmer part no. N101-1206 (You may use this cable for the Autosystem (XL) and Clarus 500, also.) Note: If you do not have the appropriate cable, verify that
|
Device Settings: |
Before installing the sampler in Chromeleon, verify that the following settings were made:
Tip: Dionex recommends selecting the pressure unit in such a way that the setting on the touch screen matches the setting in Chromeleon (set on the Options tab page in the Server Configuration program). |
Restrictions: |
Chromeleon does not support the following options:
|
Software Installation
Install the Device Driver in the Chromeleon Server Configuration. (For more information about the program, refer to Software Installation and Communication The Server Configuration Program.)
Start the sampler and the Chromeleon server.
Start the Server Configuration program.
Add the TurboMatrix Headspace sampler to the timebase. Select Add Device on the Edit or context menu, and then select Perkin Elmer from the Manufacturers list box and TurboMatrix Headspace Sampler from the Devices list box.
A wizard guides you through the initial installation. The following settings are required:
On the General page, select a free COM port and click Next.
On the Options tab page, select the device type.
Initial device configuration is now complete. Click Finish to exit he installation wizard. You can reopen the sampler configuration later again. Chromeleon then provides two additional tab pages:
On the Relays tab page, select the relays to be used.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the sampler, connect the PE_TM.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
The User Help section provides information about the device-specific commands as well as tips for certain applications; refer to Device Control: Commands and Tips for Third-Party Devices Perkin Elmer TurboMatrix Headspace Sampler.
Troubleshooting
If you have special problems, please contact Dionex Service.
Further Information
Tip:
For more installation details and for information about the different applications, refer to the manual for the instrument.