Installing and Controlling Third-Party
Devices
Bio-Rad AS 100 Autosampler: Installation
Device Type: |
Autosampler |
Device Driver: |
BIO-RAD: AS 100 Sampler |
Supported Hardware Options:
|
No special options available. |
What is required?
License: |
Timebase Class 2 |
Connection: |
RS-232
cable (see |
Hardware |
A free RS-232 port. |
BIORAD_AS100.pan |
Hardware Installation
Software Installation
Install the Device Driver in the
Chromeleon Server Configuration. (For more information about the program,
refer to Software Installation and Communication
The
Server Configuration
Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the autosampler to the timebase. Select Add Device on the Edit or context menu, and then select BIO-RAD from the Manufacturers list box and AS 100 Sampler from the Devices list box.
The following settings are required:
On the General tab page, select a free COM port.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the Bio-Rad AS 100 autosampler, connect the BIORAD_AS100.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
Example for a typical AS 100 Program
-0.1 Flow = 1.000
Sampler.Temperature.Nominal = 5
; make sure that the sampler is ready for inject
Sampler.Flush
0.000 ; wait until sampler becomes ready for inject
Wait Sampler.ReadyForInject
UV.Autozero
Inject
UV_VIS_1.AcqOn
10.000 UV_VIS_1.AcqOff
End
Further Information
Tip:
For more installation details, refer to the manual for the instrument.
For an overview of manufacturers whose instruments
can be controlled by Chromeleon, refer to Installing
and Controlling Third-Party Devices.