Maximum Height Reject

 

Location:

Detection tab page
Report Variable (Detection Parameters)

Type:

Fixed point number

Dimension:

Signal unit, such as mAU

Value Range:

0.000 ... 1,000,000.000

Default:

The default setting is Off; i.e., all peaks are integrated.

Related Parameters:

Minimum Height

Maximum Peak Height

Maximum Area Reject

Description:

Select the Maximum Height Reject parameter to limit the number of integrated peaks. This parameter is very similar to Minimum Height.

Function:

The parameter defines the maximum peak height up to which a peak is rejected. Not all peaks with a peak height below the defined value are integrated. (Determination of the height is always relatively to the baseline.)

 Tip:

Contrary to Minimum Height, the Maximum Height Reject parameter does not affect the baseline course.

 

For information about how to use individual detection parameters, refer to Integrating Chromatograms and Identifying Peaks  Defining Detection Parameters.