Maximum Area Reject
Location: |
Detection tab page |
Type: |
Fixed point number |
Dimension: |
Signal unit * Minute |
Value Range: |
0.000 ... 1,000,000.000 |
Default: |
The default setting is Off; i.e., all peaks are integrated. |
Related Parameters: |
|
Description: |
Select the Maximum Area Reject parameter to limit the number of integrated peaks. This parameter is very similar to Minimum Area. |
Function: |
The parameter defines the maximum peak area up to which a peak is rejected. No peaks with a peak area below the defined value are integrated. |
Tip: |
Contrary to the Minimum Area the Maximum Area Reject parameter does not affect the baseline course. |
For information about how to use individual detection parameters, refer to Integrating Chromatograms and Identifying Peaks Defining Detection Parameters.