Method (QNT Method)
Type: |
File name |
Dimension: |
--- |
Value Range: |
--- |
Default: |
Name of QNT File |
Related Parameters: |
Program (PGM File) |
Description: |
The Method column (in the Browser) contains the name of the quantification method (see The QNT Editor The Quantification Method (QNT Editor)). The quantification method includes all parameters that are used for evaluating a peak or the entire chromatogram. |
Function: |
The QNT Method serves as the basis for calculation for sample evaluation. It includes: All parameters required for (qualitative) peak identification and for converting the determined peak areas into amount or concentration values (quantitatively) All parameters regarding, for example, peak recognition and peak area evaluation All parameters regarding the type and performance of a calibration Information about baseline subtraction Peak Tracking Parameters for comparing spectra to library or sequence spectra Spectra Library Screening Parameters for spectra library screening Parameters for performing the System Suitability Test MS Parameters for Mass Spectra evaluation |
Tip: |
Normally it is not necessary to include the QNT Method in the sample list before the analysis starts. However, if you want to perform a System Suitability Test, make sure to enter the QNT File into the sample list before starting the analysis. Otherwise, the batch cannot be aborted in case of Fail Action - Abort Batch because the SST will not yet be performed during the batch run! |