Detector Control
Displaying the Signal Parameters of a Mass Channel
Different channel types are available for Mass Spectrometers:
TICF channels (TICF 01 to TICF 09 for the MSQ or TICF_1 to TICF_4 for the aQa: data acquisition in Full-Scan mode)
SIM channels (SIM_01 to SIM_32: data acquisition in SIM mode)
TIC channel (total of all TICF or SIM channels)
MS_01 to MS_32 channels (data acquisition in full-scan mode)
Almost no parameters for mass spectrometer channels can be changed from the Control Panel. Select Axis/Decoration on the context menu of the signal plot to display the following parameters for the channels of a mass spectrometer:
Parameter |
Description |
Comment |
Filter |
Filter of the MS method (METH file). |
No input possible, not even in the F8 dialog box. |
FilterIndex |
Number of the Filter of the MS method. |
Only for the MS_01 to MS_32 channels (press F8): Select the number of the desired filter (1-9 for the MSQ or 1-4 for the aQa): |
Min/MaxMass |
Limits of the mass range that is acquired in this channel. |
Only for MS channels MS_01 to MS_32 (press the F8 key): Enter the lower and/or upper mass range limit. |
Polarity |
Ionization polarity of the mass spectrometer. |
No input possible, not even in the F8 dialog box. |
Source Voltage |
Voltage of the ion source of the mass spectrometer. |
No input possible, not even in the F8 dialog box. |
Press the F8 key or select Commands on the Control menu to display the following parameters:
Parameter |
Description |
Comment |
Delta |
Difference between the current signal value and the signal value one second before. |
The value cannot be changed. |
Retention |
Indicates the current retention time. |
The value cannot be changed. |
Signal |
Value indicates the current value of a Signal. UpperLimit and LowerLimit define the signal acquisition limits. |
If signal acquisition limits are not met, the running batch is stopped and the emergency program is started. |