Detector Control
Recording Fluorescence Spectra (FLD-3x00RS): Program Examples
The Dionex Ultimate3000 FLD-3x00(RS) fluorescence detectors allow you to record the fluorescence spectra of substances used in an analysis. Spectrum scans can be performed manually by executing a command (for more information, refer to Recording Fluorescence Spectra) or in a Program. Below are several examples for spectrum scanning in a program.
Tip:
Record the baseline spectrum before the scan, at a time when no peak occurs. It is stored in the detector and automatically subtracted from the recorded spectra. It is not mandatory to perform a baseline scan prior to every excitation/emission/synchro scan. However, you must record a new baseline spectrum before the scan whenever you change the scan settings. The ClearBaseline command deletes the current baseline spectrum.
Emission scan with preceding baseline scan
Run an emission scan to record the emission spectrum of an analyte. The excitation wavelength remains constant while the emission wavelength is scanned over a defined wavelength range. The spectrum maximum indicates the optimum emission wavelength.
The program example below shows a simple emission scan program, with a preceding baseline scan. Adjust the parameters as needed for your application.
0.100 |
ScanStartExWavelength = |
250.0 [nm] ;defines the constant excitation wavelength |
|
ScanStartEmWavelength = |
340.0 [nm] ;defines the lower emission scan limit |
|
ScanEndEmWavelength = |
480.0 [nm] ;defines the upper emission scan limit |
|
ScanPMT = |
Pmt1 ;defines which PMT is used during the scan (only for detectors with second PMT) |
|
ScanSpeed = |
Fast ;defines the speed of any following scan |
|
ScanSensitivity = |
5 ;defines the sensitivity of the PMT during the scan |
|
ScanFilterWheel = |
280nm ;selects the position of the filter wheel (only for FLD-3400RS) |
|
ScanSignalType = |
Standard ;determines if the recorded signal will be corrected regarding the xenon lamp spectrum. Standard will provide the best signal-to-noise ratio in method development. |
|
BaselineBehavior = |
Append ;defines that the baseline is appended to the previous signal after a wavelength, sensitivity, PMT, or filter switch |
|
BaselineScanEmission |
;executes a baseline scan |
0.500 |
ScanEmission |
;executes an emission scan |
Excitation scan with preceding baseline scan
The program example below shows an example of an excitation scan with preceding baseline scan. Note that some commands are similar to an emission scan. Adjust the parameters as needed for your application.
0.100 |
ScanStartEmWavelength = |
480.0 [nm] ;defines the constant emission wavelength |
|
ScanStartExWavelength = |
250.0 [nm] ;defines the lower excitation scan limit |
|
ScanEndExWavelength = |
450.0 [nm] ;defines the upper excitation scan limit |
|
ScanPMT = |
Pmt1 ;defines which PMT is used during the scan (only for detectors with second PMT) |
|
ScanSpeed = |
Fast ;defines the speed of any following scan |
|
ScanSensitivity = |
5 ;defines the sensitivity of the PMT during the scan |
|
ScanFilterWheel = |
280nm ;selects the position of the filter wheel (only for FLD-3400RS) |
|
ScanSignalType = |
Standard ;determines if the recorded signal will be corrected regarding the xenon lamp spectrum. Standard will provide the best signal-to-noise ratio in method development. |
|
BaselineBehavior = |
Append ;defines that the baseline is appended to the previous signal after a wavelength, sensitivity, PMT, or filter switch |
|
BaselineScanExcitation |
;executes a baseline scan |
0.500 |
ScanExcitation |
;executes an excitation scan |
Synchro scan with preceding baseline scan
The program example below is an example of a synchro scan with preceding baseline scan. A synchro scan alternates both the excitation and the emission wavelengths across a defined wavelength range. During the scan, the offset between excitation and emission is kept constant. This scan type is useful to determine a functioning wavelength pair. Thereby, the maximum of the synchro spectrum indicates the best wavelength combination at the given offset. Conversely, the synchro spectrum does typically not indicate the optimum wavelength combination. Adjust the parameters as needed for your application.
0.100 |
ScanStartExWavelength = |
250.0 [nm] ;defines the lower excitation scan limit |
|
ScanEndExWavelength = |
450.0 [nm] ;defines the upper excitation scan limit |
|
ScanSyncOffset = |
50.0 [nm] ;defines the offset between Ex and Em wavelength. The Em range starts at StartExWavelength + 50 nm (= 300 nm) and ends at EndExWavelegnth + 50 nm (= 500 nm). |
|
ScanPMT = |
Pmt1 ;defines which PMT is used during the scan (only for detectors with second PMT) |
|
ScanSpeed = |
Fast ;defines the speed of any following scan |
|
ScanSensitivity = |
5 ;defines the sensitivity of the PMT during the scan |
|
ScanFilterWheel = |
280nm ;selects the position of the filter wheel (only for FLD-3400RS) |
|
ScanSignalType = |
Standard ;determines if the recorded signal will be corrected regarding the xenon lamp spectrum. Standard will provide the best signal-to-noise ratio in method development. |
|
BaselineBehavior = |
Append ;defines that the baseline is appended to the previous signal after a wavelength, sensitivity, PMT, or filter switch |
|
BaselineScanSynchronous |
;executes a baseline scan |
0.500 |
ScanSynchronous |
;executes an synchro scan |
Emission scan using a trigger
The trigger example below shows a Trigger that starts a scan at a signal change of 250.000 counts/second or when reaching 2.500.000 counts. The trigger automatically reduces the flow to zero, defines the scan conditions, performs the scan and then increases the flow rate to the initial value. The trigger is activated each time the conditions become true. The trigger can be used to record spectra for any peaks independent of the retention time.
You can add this trigger example to your program by adjusting the values as needed.
;Trigger Control |
||
1.000 |
Trigger PEAK (Emission_1.Delta > 250000) OR (Emission_1 > 2500000) ;activates the trigger when the signal increases for more than 250,000 counts/second or total signal level exceeds 2,500,000 counts |
|
|
MaximumFlowRampDown = |
Infinite ;defines the speed with which the pump reduces the flow rate |
|
Flow = |
0.000 [ml/min] ;sets the pump flow to zero to park the peak in flow cell during scan |
|
FLD.BaselineBehavior = |
Append ;defines that after a wavelength, sensitivity, PMT, or filter switch, the baseline is appended to the previous signal |
|
FLD.ScanSpeed = |
Fast ;defines the emission scan speed |
|
FLD.ScanStartExWavelength = |
255.0 [nm] ;defines the constant excitation wavelength |
|
FLD.ScanStartEmWavelength = |
275.0 [nm] ;defines the lower emission scan limit |
|
FLD.ScanEndEmWavelength = |
600.0 [nm] ;defines the upper emission scan limit |
|
FLD.ScanSignalType = |
Standard ;determines if the recorded signal will be corrected regarding the xenon lamp spectrum. Standard will provide the best signal-to-noise ratio in method development. |
|
FLD.ScanEmission |
;starts the emission spectrum scan |
|
Wait FLD.Ready |
;highly important - restarts pump flow only AFTER scan has finished |
|
MaximumFlowRampUp = |
10 [ml/minē] ;defines the speed with which the pump increases the flow rate ;outside trigger control you can set MaximumFlowRampUp to a different value (e.g., to slowly increase the flow at system startup) |
|
Flow = |
1.000 [ml/min];sets the pump flow to the initial value. |
|
EndTrigger |
|
;End of Trigger |
|
Excitation scan using a trigger
The trigger example below shows an excitation scan with a trigger as described above. You can add this trigger example to your program by adjusting the values as needed.
;Trigger Control |
||
1.000 |
Trigger PEAK (Emission_1.Delta > 250000) OR (Emission_1 > 2500000) ;activates the trigger when the signal increases for more than 250,000 counts/second or total signal level exceeds 2,500,000 counts |
|
|
MaximumFlowRampDown = |
Infinite ;defines the speed with which the pump reduces the flow rate |
|
Flow = |
0.000 [ml/min] ;sets the pump flow to zero to park the peak in flow cell during scan |
|
FLD.BaselineBehavior = |
Append ;defines that after a wavelength, sensitivity, PMT, or filter switch, the baseline is appended to the previous signal |
|
FLD.ScanSpeed = |
Fast ;defines the emission scan speed |
|
FLD.ScanStartEmWavelength = |
500.0 [nm] ;defines the constant emission wavelength |
|
FLD.ScanStartExWavelength = |
250.0 [nm] ;defines the lower excitation scan limit |
|
FLD.ScanEndExWavelength = |
450.0 [nm] ;defines the upper excitation scan limit |
|
FLD.ScanSignalType = |
Standard ;determines if the recorded signal will be corrected regarding the xenon lamp spectrum. Standard will provide the best signal-to-noise ratio in method development. |
|
FLD.ScanExcitation |
;starts the excitation spectrum scan |
|
Wait FLD.Ready |
;highly important - restarts pump flow only AFTER scan has finished |
|
MaximumFlowRampUp = |
10 [ml/minē] ;defines the speed with which the pump increases the flow rate ;outside trigger control you can set MaximumFlowRampUp to a different value (e.g., to slowly increase the flow at system startup) |
|
Flow = |
1.000 [ml/min];sets the pump flow to the initial value. |
|
EndTrigger |
|
;End of Trigger |
|
Synchro scan using a trigger
The trigger example below shows a synchro scan with a trigger as described above. You can add this trigger example to your program by adjusting the values as needed.
;Trigger Control |
||
1.000 |
Trigger PEAK (Emission_1.Delta > 250000) OR (Emission_1 > 2500000) ;activates the trigger when the signal increases for more than 250,000 counts/second or total signal level exceeds 2,500,000 counts |
|
|
MaximumFlowRampDown = |
Infinite ;defines the speed with which the pump reduces the flow rate |
|
Flow = |
0.000 [ml/min] ;sets the pump flow to zero to park the peak in flow cell during scan |
|
FLD.BaselineBehavior = |
Append ;defines that after a wavelength, sensitivity, PMT, or filter switch, the baseline is appended to the previous signal |
|
FLD.ScanSpeed = |
Fast ;defines the emission scan speed |
|
FLD.ScanStartExWavelength = |
250.0 [nm] ;defines the lower excitation scan limit |
|
FLD.ScanEndExWavelength = |
450.0 [nm] ;defines the upper excitation scan limit |
|
FLD.ScanSyncOffset = |
50.0 [nm] ;defines the offset between Ex and Em wavelength. The Em range starts at StartExWavelength + 50 nm (= 300 nm) and ends at EndExWavelegnth + 50 nm (= 500 nm). |
|
FLD.ScanSignalType = |
Standard ;determines if the recorded signal will be corrected regarding the xenon lamp spectrum. Standard will provide the best signal-to-noise ratio in method development. |
|
FLD.ScanSynchronous |
;starts the synchronous spectrum scan |
|
Wait FLD.Ready |
;highly important - restarts pump flow only AFTER scan has finished |
|
MaximumFlowRampUp = |
10 [ml/minē] ;defines the speed with which the pump increases the flow rate ;outside trigger control you can set MaximumFlowRampUp to a different value (e.g., to slowly increase the flow at system startup) |
|
Flow = |
1.000 [ml/min];sets the pump flow to the initial value. |
|
EndTrigger |
|
;End of Trigger |
|
For a list of all fluorescence detector commands, refer to Dionex UltiMate 3000 Series Fluorescence Detectors.