Commands and Tips for Third-Party Devices
Waters Alliance 2790/2795: HPLC System

In addition to the standard commands (refer to Practical Tips for Device Control  General Device Commands), the Waters 2790/2795 Separation Module device driver supports the following commands and properties for the HPLC System (please note that the display Filter level determines which commands and properties are displayed):

Property

Values

Description

ChartOut²

³ GradA, GradB, GradC, GradD, FlowRate, SampleLoopPressure, SystemPressure, PrimaryHeadPressure, VacuumLevel, SampleTemperature, ColumnTemperature

Sets the type of the analog signal that is transmitted from the Chart Out output.

The available options depend on the system configuration. The following options are available only if the corresponding module is installed:

Option

Module

VacuumLevel

Degasser

SampleTemperature

Sample Heater

ColumnTemperature

Column Heater

LogStatus²

On, Off

If the property is set to On, each change of the instrument's status is logged in the Audit Trail. (The default setting is Off. The property is visible only on the Advanced level.)

PrecolumnVolume²

0.0 to 10,000.0

Specifies the solvent volume in [µl] that is pumped through the column before injection is started. (The default setting is 0.0 µl. The property is visible only on the Advanced level.)

Status

(80 characters)

Indicates the status of the HPLC system.

Vendor

 

Indicates the manufacturer (here: Waters).

 Tips: 

² It is not possible to change the property during injection or during a gradient run.

³ The signal condition depends on the selected ChartOut option:

Signal

Parameter at 0 mV (min)

Parameter at 2000 mV (max)

ColumnTemperature

20°C

60°C

FlowRate

0.000 ml/min

10.000 ml/min

GradA - GradD

0.0%

100.0%

PrimaryHeadPressure

-3.45 bar (-50 psi)

380 bar (5500 psi)

SampleLoopPressure

-3.45 bar (-50 psi)

380 bar (5500 psi)

SampleTemperature

4°C

40°C

SystemPressure

-3.45 bar (-50 psi)

380 bar (5500 psi)

VacuumLevel

-0.0 bar

0.34 bar (4.91 psi)

 

Commands

 

Command

Description

StopOperation

Stops a previously started operation (e.g., by the SysPrep or WetPrime commands (see below) or by the PrimeSyringe command, see  Waters Alliance 2790/2795 HPLC Modules: Injector). (The command is visible only on the Advanced level.)

SysPrep

Performs system preparation including priming the fluid components of the systems, rinsing the syringe, and washing the needle. (The command is visible only on the Advanced level.)

WetPrime

Primes the fluid components of the system with the currently selected flow rate and for 100.0 minutes. Use the StopOperation command to stop priming. (The command is visible only on the Advanced level.)

 

For information about how to install the Waters Alliance 2790 and 2795 HPLC modules, refer to Hardware Installation  Waters Alliance 2790/2795 HPLC Modules: Installation in the Administrator Help section.