Commands and Tips for Third-Party Devices
Waters Alliance 2790/2795: Injector
For information about the standard autosampler commands, refer to Control Dionex Autosampler.
In addition, the Waters 2790/2795 Separation Module device driver supports the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
Property |
Min |
Max |
Description |
|
CheckPlateHeight² |
Off |
On |
When set to On, the needle position sensor determines the plate height before the first injection is performed from this plate. The value is checked against the range of values allowed for this plate type. (The default setting is Off. The property is visible only on the Advanced level.) |
|
FullLoopOverfill² |
1.0% |
99.9% |
Only in InjectMode = FullLoop (see below). |
|
InjectWaitTime |
0 |
1E+7 |
Indicates the time in seconds between the inject command from Chromeleon and the inject response from the autosampler to Chromeleon. (The property is read-only.) |
|
InjPortWashTime² |
0, 1, 2, ... 9999 |
Specifies the time in seconds for inject port washing. (The default setting is 6. The property is visible only on the Advanced level.) |
||
InjectMode² |
PartialLoop; FullLoop |
Determines the injection mode. (The default setting is FullLoop.) |
||
FullLoop |
The entire loop volume is delivered to the column. |
|||
PartialLoop |
Only part of the loop volume is delivered to the column. |
|||
LoopPressure |
-3.45 |
380 |
Indicates the pressure (in bar) in the sample loop (The property is read-only and visible only on the Expert level.) |
|
NeedleWashTime² |
0, 1, 2, ... 9999 |
Specifies in seconds how long the exterior of the needle is washed. (The default setting is 15. The property is visible only on the Advanced level.) |
||
PrepDelay³ |
0.0 |
Loop |
Determines in seconds how long it may take until the loop is completely filled. (The default setting is 0.0. The property is visible only on the Advanced level.) |
|
PrepSubject³ |
Vial positions Sample_Vial, Air, Sample_Vial ± number (relative addressing) |
Determines which vial is used for Draw, Dispense, and Mix operations. If set to Air, the needle first draws air, and then dispenses the air into the needle port. (The default setting is 15. The property is visible only on the Advanced level.) |
||
PrepVolume³ |
0, 1, 2, ... 9999 |
Determines the volume [µl] to be used for the Draw, Dispense, and Mix commands. (The default setting is 10.0. The property is visible only on the Advanced level.) |
||
|
Primes the syringe by flushing it. Tip: After priming the syringe, always perform a Wait Sampler.Ready command. |
|||
PurgeVolume³ |
0, 1, 2, ... 9999 |
Determines the volume [µl] to be used for purging. (The default setting is 100. The property is visible only on the Advanced level.) |
||
NotReady |
Ready |
State of the autosampler (depends on the state of the sample temperature control if WaitForTemperature = Yes). |
||
SampleHeight² |
0.0 mm |
20.0 mm |
Indicates the distance of the needle tip from the bottom of the sample vial (default: 0.0 mm). |
|
SampleOverlapped² |
Off |
On |
When set to On, the current sample will be prepared while the previous sample is still analysed. (The default setting is Off. The property is visible only on the Advanced level.) |
|
SeekWellBottom² |
Off |
On |
Can be enabled only if SampleHeight = 0.0 (see above). When set to On, the needle detects the well bottom for a specified plate before the first injection is performed. This value is stored for the subsequent injections from this plate. When a different plate is used, the measurement is performed again. (The default setting is Off. The property is visible only on the Advanced level.) |
|
State |
Off |
On |
Indicates On when the autosampler has injected a sample. (The property is read-only.) |
|
SyringeSpeed², |
Fast (5.0 µl/s) |
Defines the drawing speed of the syringe. The speed depends on the installed syringe type. The values correspond to a standard 250 µl syringe. |
||
WaitForTemperature² |
No |
Yes |
Specifies whether the autosampler waits for the sample temperature control. |
|
|
Flushes the autosampler. Tip: After priming the syringe, always perform a Wait Sampler.Ready command. |
|||
WashCycle² |
0, 1, 2, ... 9 |
Sets the number of wash cycles to be performed. (The default setting is 1. The property is visible only on the Advanced level.) |
||
WashFrequency² |
Never, EveryInject, EveryWell |
Determines when a wash cycle shall be performed. The following options are available: Never, EveryInject, EveryWell. (The default setting is EveryInject. The property is visible only on the Advanced level.) |
||
WashSequence² |
PurgeWash, PurgeWashPump |
Determines the wash sequence for the needle. (The default setting is PurgeWash. The property is visible only on the Advanced level.) |
Tips:
² It is not possible to change the setting during an injection, during gradient runs, or at negative retention times.
³ It is not possible to change the setting during sample analysis.
Note:
As described in the instrument's manual, negative pressure readings are possible for the sample loop pressure (LoopPressure property).
Injector Commands
Command |
Description |
Draw³ |
Draws the PrepVolume from the PrepSubject (see above). (The command is available only on Advanced level.) Tips: Up to 16 Draw commands are allowed per injection. Always use a Mix command after a sequence of Draw commands. |
LoopOffline³ |
Switches the sample injection valve after the injection. Thus, the loop is no longer part of the flow path. Preparation of the next sample can be started. (The property is visible only on the Advanced level.) Tip: Make sure to issue the command only after the previous sample has completely left the loop. |
Mix³ |
Starts sample preparation. (The property is visible only on the Advanced level.) Tips: No further Draw commands are allowed after a Mix command. Always use a Mix command after a sequence of Draw commands. |
PrimeSyringe |
Fills the syringe with the purge solvent. (The property is visible only on the Advanced level.) Tip: The command uses the values set for PurgeVolume and WashCycle. |
Tip:
³ It is not possible to change the setting during sample analysis.
For information about how to install the Waters Alliance 2790 and 2795 HPLC modules, refer to Hardware Installation Waters Alliance 2790/2795 HPLC Modules: Installation in the Administrator Help section.