Commands and Tips for Third-Party Devices
Waters 2487 UV Detector
Tip:
Before data acquisition can start, the detector needs some time to load the parameter settings. That is why a Wait command is required before data acquisition is started.
In addition to the standard commands, the Waters 2487 UV Detector provides the following commands (please note that the display Filter level determines which commands are displayed):
Command |
Min |
Max |
Default |
Description |
Data_Collection_Rate |
1 Hz |
10 Hz |
1 Hz |
Indicates the data collection rate for the channels. |
Lamp |
Off |
On |
|
Turns the lamp on or off. |
Commands and Properties of UV Channels
Property |
Min |
Max |
Default |
Description |
AutozeroOn |
On |
Off |
Off |
If the property is enabled, autozeroing is performed whenever this is required by an event. |
AutozeroOn |
On |
Off |
On |
If the property is enabled, autozeroing is performed each time the wavelength is changed. Note: Autozero is performed for all channels, even if only the wavelength of a single channel is changed. |
Off |
On |
On |
Allows signal averaging. |
|
FilterTime |
0.0 s |
99.0 s |
1.0 s |
Time constant for the selected noise filter. |
FilterType |
None |
RC |
Hamming |
Filter type used to reduce analog and digital signal noise. |
Signal. |
-10^10 mAU |
10^10 mAU |
0 |
Indicates the upper and lower limit for the signal. |
Auto, 0.01 s |
4.80 s |
Auto |
Interval between two data points. |
|
190 nm |
700 nm |
UV_VIS_1: 254.0 nm |
Wavelength at which the detector measures the signal (see the Troubleshooting section). |
Tips:
² It is only possible to change the Filtertype, AutozeroOnEventIn, and AutozeroOnWlChange properties once in the PGM File during negative retention times.
³ It is not possible to change the time constant for the selected filter during data acquisition.
For information about how to install the Waters 2487 UV detector, refer to Hardware Installation Waters 2487 UV Detector: Installation in the Administrator Help section.