Commands and Tips for Third-Party Devices
Varian CP-3800 GC: General
In addition to the General Device Commands, the Varian CP-3800 GC supports the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
Property |
Range/Value |
Description |
|
AtmosphericPressure˛ |
0.000 - 2.067 bar |
Indicates the atmospheric pressure |
|
Cryo |
On/Off |
(If cryo cooling is installed) Enables or disables cyro cooling. |
|
CryoHasTimeout˛ |
Yes, No |
(If cryo cooling is installed) = Yes if the temperature zone did not reach the target temperature, despite Cryo. |
|
CryoTimeout |
0.01 - 999.99 min |
(If cryo cooling is installed) Time that the GC waits before turning off the cooling if no sample has been injected. |
|
CryoThreshold |
30-450°C |
(If cryo cooling is installed) Temperature above which the GC shall use cryo for cooling. |
|
EndRunTime˛ |
0.01 - 999.99 min |
Actual run time of the current program. |
|
Fault˛ |
Fault, OK |
Indicates whether a fault occurred in the GC*. |
|
OvenFault˛ |
Fault, OK |
Indicates whether a fault occurred in the column oven. |
|
OvenPower˛ |
On, Off |
Indicates whether the column oven is on. |
|
OvenReady˛ |
Ready, NotReady |
Indicates whether the column oven has reached the target temperature and is ready. |
|
ProtocolVersion |
n/a |
n/a |
Protocol version of the instrument (must be '7'). |
RunState |
NotReady, Equilibrating, Stabilizing, Monitoring, Ready, Running |
Indicates the status of the GC. |
|
RunTime |
0.01 - 999.99 min |
Indicates the actual retention time during sample processing. |
|
StabilizationTime˛ |
0.00 - 10.00 min |
Time that the GC waits after reaching the target temperature. |
|
Temperature |
|
|
Properties related to the column oven temperature. |
Temperature. |
-99°C - 450°C |
Current temperature. |
|
Temperature. |
-99°C - UpperLimit |
Sets the nominal temperature. |
|
Temperature. |
50 - 450°C |
The system aborts the sample batch and starts emergency handling if the temperature exceeds the upper limit. |
Note:
˛ These properties are read-only.
Control with or without a non-controlled autosampler:
To synchronize the injection command of the GC with the start of data acquisition, enter the following command instead of the Inject command:
Wait GC.RunState=Running
This applies to injections via a headspace sampler or a valve.
For installation details, refer to Hardware Installation: Installing and Controlling Third-Party Devices Varian CP-3800 GC: Installation in the Administrator Help section.