Commands and Tips for Third-Party
Devices
Varian 3400 and Varian 3600 GCs: Detectors
The Varian 3400 and 3600 GC device drivers support the following commands and properties for the different detectors (please note that the display Filter level determines which commands and properties are displayed):
General Detector Properties
Property |
Min |
Max |
Default |
Description |
Attenuation |
1 |
8 |
1 |
Sets the attenuation of the detector signal |
Baseline |
-1315.80 |
1315.80 |
N/a |
Indicates the actual baseline value. |
TempCtrl |
Off |
On |
On |
Turns the detector heater on or off. |
Temperature.Nominal |
² |
² |
² |
Sets the nominal temperature for the detector heater. |
Temperature.UpperLimit |
² |
² |
² |
Sets the maximum temperature for the detector heater. |
Temperature.Value |
-99°C |
420°C |
50°C |
Indicates the actual temperature of the detector heater (read-only). |
Type |
TCD, FID, TSD, or ECD |
N/a |
Indicates which type of detector is installed. |
² The values depend on the device settings.
Command |
Description |
Sets the current detector signal to 0. |
Thermal Conductivity Detector (TCD)
Property |
Min |
Max |
Default |
Description |
FilamentTemp |
50°C |
390°C |
50°C |
Sets the temperature of the TCD filament. |
Polarity |
Negative |
Positive |
Positive |
Sets the polarity of the detector signal. |
Range |
5 |
500 |
500 |
Range for the output signal: 5, 50, 500. |
Flame Ionization Detector (FID)
Property |
Min |
Max |
Default |
Description |
Range |
10^8 |
10^12 |
10^8 |
Range for the output signal. |
Thermionic Specific Detector (TSD)
Property |
Min |
Max |
Default |
Description |
BeadCurrent |
2400 |
3800 |
2400 |
Indicates the current bead power [mA]. |
BeadPower |
Off |
On |
On |
Turns the TSD bead power on or off. |
Range |
10^8 |
10^12 |
10^8 |
Range for the output signal. |
Electron Capture Detector (ECD)
Property |
Min |
Max |
Default |
Description |
Range |
1 |
10 |
1 |
Range for the output signal (allowed values: 1 and 10). |
Also, refer to Varian 3400 3600 GC Detectors: Different Names in Firmware and Chromeleon.