Commands and Tips for Thermo Scientific Devices
Finnigan/TSP UV3000 Detector
The following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
Property |
Min |
Max |
Default |
Description |
AcquisitionMode |
Single |
Multi |
N/A |
Changes the data acquisition mode. |
Connected |
Discon- |
Connected |
N/A |
Indicates whether the detector is connected. |
UV_Lamp |
Off |
On |
N/A |
Turns the deuterium lamp on or off (l = 190 - 365 nm) |
UV_Lamp_ |
Intensity_ |
Intensity_ |
N/A |
Indicates whether the deuterium lamp intensity is too low or OK (read-only) |
Visible_Lamp |
Off |
On |
N/A |
Turns the tungsten lamp on or off (l = 366 - 800 nm) |
Visible_Lamp_ |
Intensity_ |
Intensity_ |
N/A |
Indicates whether the tungsten lamp intensity is too low or OK (read-only) |
Channel-specific Commands and Properties
Property |
Min |
Max |
Default |
Description |
Response |
0.0 |
9.9 |
1.0 |
Time constant. Indicates the time needed by the detector to reach 63% of the full-scale deflection. |
190 nm |
800 nm |
254 nm |
Wavelength at which the detector measures the signal. |
All other channel properties and commands correspond to the standard detector commands (refer to Practical Tips for Device Control Detector Commands).
Notes:
The Range property provided by the TSP UV3000 (analog data acquisition) driver is not required for digital data acquisition; thus, it is not available with the TSP UV3000 (digital data acquisition) driver.
Due to extensive communication during the Connect procedure, it may take up to 1 min until the instrument is finally connected. The corresponding message is logged in the Audit Trail.
Tip:
The device driver does not support the Scan mode of the detector. Therefore, it is not possible to acquire single spectra.