Detector Control
Detector Commands
Modifying Signal Parameters
0.000 |
Channelname.Average |
=On / Off |
0.000 |
Channelname.Bandwidth |
=Value |
0.000 |
Channelname.Wavelength |
=Value |
0.000 |
Channelname.Step |
=Value [s] / Auto |
Tip:
Select the Step in such a way that 20 to 50 data points are placed on the narrowest relevant peak in your chromatogram. If you acquire more data points than necessary, more storage space is consumed and data processing slows. In addition, integration might be incorrect, especially when baseline noise is high. This may also happen when using the Step =Auto option.
Note:
To implement Wavelength Switching, define a new wavelength for a channel at the switch time.
Switching Detector Lamp
0.000 |
Devicename.Lamp |
=On / Off |
Switching Detector Lamps On/Off
AD20 Detector
0.000 |
Devicename.UV_Lamp |
=Low / High / Off |
0.000 |
Devicename.Visible_Lamp |
=Low / High / Off |
AD25, PDA, PDA-100, and PDA-3000 Detectors
0.000 |
Devicename.UV_Lamp |
=On / Off |
0.000 |
Devicename.Visible_Lamp |
=On / Off |
Selecting the Detection Mode
ED40, ED50, and ED50A Detectors
0.000 |
Devicename.Mode |
=Conductivity / IntAmp / DCAmp |
ICS-3000 ED and ICS-5000 ED Detectors
0.000 |
Devicename.Mode |
=IntAmp / DCAmp / Cyclic |
ICS-4000 ED Detector
0.000 |
Devicename.Mode |
=IntAmp / DCAmp |
Turning the Amperometry Cell On/Off
ED40, ED50, and ED50A Detectors in Amperometry Mode
0.000 |
Devicename.Cell |
=On / Off |
ICS-3000 ED, ICS-4000 ED, and ICS-5000 ED Detectors
0.000 |
Devicename.CellControl |
=On / Off |
Selecting the Reference Electrode
ICS-3000 ED, ICS-4000 ED, ICS-5000 ED, and ED40/ED50/ED50A Detectors in Amperometry Mode
0.000 |
Devicename.Electrode |
=pH / AgCl / PdH |
Note:
The PdH reference electrode is available with the ICS-4000 ED and ICS-5000 ED only.