Detector Control
Detector Commands

Modifying Signal Parameters

 

0.000

Channelname.Average

=On / Off

0.000

Channelname.Bandwidth

=Value

0.000

Channelname.Wavelength

=Value

0.000

Channelname.Step

=Value [s] / Auto

 

 Tip: 

Select the Step in such a way that 20 to 50 data points are placed on the narrowest relevant peak in your chromatogram. If you acquire more data points than necessary, more storage space is consumed  and data processing slows. In addition, integration might be incorrect, especially when baseline noise is high. This may also happen when using the Step =Auto option.

 Note:

To implement Wavelength Switching, define a new wavelength for a channel at the switch time.

Switching Detector Lamp

 

0.000

Devicename.Lamp

=On / Off

 

Switching Detector Lamps On/Off

AD20 Detector

0.000

Devicename.UV_Lamp

=Low / High / Off

0.000

Devicename.Visible_Lamp

=Low / High / Off

 

AD25, PDA, PDA-100, and PDA-3000 Detectors

0.000

Devicename.UV_Lamp

=On / Off

0.000

Devicename.Visible_Lamp

=On / Off

 

Selecting the Detection Mode

ED40, ED50, and ED50A Detectors

0.000

Devicename.Mode

=Conductivity / IntAmp / DCAmp

 

ICS-3000 ED and ICS-5000 ED Detectors

0.000

Devicename.Mode

=IntAmp / DCAmp / Cyclic

 

ICS-4000 ED Detector

0.000

Devicename.Mode

=IntAmp / DCAmp

Turning the Amperometry Cell On/Off

ED40, ED50, and ED50A Detectors in Amperometry Mode

0.000

Devicename.Cell

=On / Off

 

ICS-3000 ED, ICS-4000 ED, and ICS-5000 ED Detectors

0.000

Devicename.CellControl

=On / Off

 

Selecting the Reference Electrode

ICS-3000 ED, ICS-4000 ED, ICS-5000 ED, and ED40/ED50/ED50A Detectors in Amperometry Mode

0.000

Devicename.Electrode

=pH / AgCl / PdH

 Note:

The PdH reference electrode is available with the ICS-4000 ED and ICS-5000 ED only.