Commands and Tips for Thermo Scientific Devices
Thermo Scientific Accela Series: Pump
The following commands and properties are available for the Accela pump (please note that the display Filter level determines which commands and properties are displayed).
Also, refer to the information in the manual shipped with the instrument.
General Commands and Properties:
Command Property |
Description |
BoardNumber |
Board number of the pump (read-only). |
Connect |
Connects the instrument to Chromeleon. |
Connected |
Indicates whether the instrument is connected to Chromeleon, that is, under computer control (also, see Connect, Disconnect). |
Disconnect |
Disconnects the instrument from Chromeleon and activates manual control from the instrument panel. Also see Connect. |
DriverID |
Unambiguous driver name (read-only). |
DriverVersion |
Driver version (read-only). |
FirmwareVersion |
Indicates the firmware version of the instrument (read-only). |
Manufacturer |
Manufacturer of the driver (read-only). |
ModelNo |
Reports the instrument's model number (read-only). |
SerialNo |
Indicates the serial number of the instrument (read-only). |
Commands and Properties for flow and pressure:
Command/Property |
Description |
Percentage of each solvent component. |
|
*.Equate |
Name of the component (A, B, C, or D). |
*.Value |
Current value [%] for components A, B, C, or D (read-only). |
Sets the flow rate [ml/min]. |
|
*.Nominal |
Reports the nominal flow. |
*.Value |
Reports the current flow (read-only). |
Pressure |
Pump pressure. |
*.LowerLimit |
Sets the lower pressure limit. The system stops the sample batch and starts emergency handling if the value is outside the set range. |
*.UpperLimit |
Sets the upper pressure limit. The system stops the sample batch and starts emergency handling if the value is outside the set range. |
*.Value |
Indicates the current pump pressure (read-only). |
(Accela Pump only) Sets the pump's pressure mode: Low for working in the high pressure range (~0-500 bar / 0-7000 PSI), or High for working in the ultra-high pressure range (~500-1000 bar / 7000-15.000 PSI). Note: You may specify a default setting for the PressureMode property on the Thermo Accela Pump Configuration page in the Chromeleon Server Configuration. Upon each Connect command, the property will be reset to the specified default. |
Commands and Properties for leak sensor and purge pump (Accela 600 and Accela 1250 only):
Note:
Support of the Accela 600 and 1250 Pumps is available starting with the SR11 Driver Update that provides extended Accela system support.
For information about how to install the instrument, refer to Hardware Installation Thermo Scientific Accela Pump: Installation.