Polymer Laboratories ELS 2100 (Ice) / Varian 380-LC (385-LC) ELS Detectors: Commands and Tips
The ELS 2100 and the ELS 380-LC/385-LC drivers support the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
General commands and tips:
Command/Property |
Description |
Autozero |
Sets the signal output to 10 mV. |
AZ_Done |
Reports the status of an autozero command (Busy or Done). |
ClearDisplayError |
Clears any error messages on the detector's display. |
Connect |
Connects the device to Chromeleon. |
Connected |
Indicates whether the detector is connected to Chromeleon, that is, under computer control (also, see Connect, Disconnect). |
Data_Collection_Rate |
Sets the rate at which Chromeleon collects digital data points from the detector (380-LC/385-LC ELS detectors only). Note: This property must be set directly on the device. |
Disconnect |
Disconnects the detector from Chromeleon and activates manual control from the panel. |
DriverID |
Unambiguous driver name. |
DriverVersion |
Indicates the driver version. |
FirmwareVersion |
Indicates the firmware version. |
Manufacturer |
Indicates the manufacturer of the driver. |
ModelNo |
Indicates the instrument's model number. |
ModelVariant |
Used by Chromeleon to distinguish between the detector variants. |
Ready |
Ready indicates that the device is ready for operation, i.e., no Autozero is started, the temperature has been reached, no error has occurred. |
Standby |
Sets the detector into the Standby mode or cancels this mode (NoStandby). In standby mode, the detector remains connected to Chromeleon. Tip: Run the detector in NoStandby mode for successful data acquisition. |
Device properties:
Command/Property |
Description |
EvaporatorTemperature |
Sets the temperature of the evaporator. Available values: Off, 25° C - 120° C or Off, 10° C - 80° C (ELS 2100 Ice). |
*.Nominal |
Actual evaporator temperature. |
*.Value |
Current evaporator temperature. |
NebuliserTemperature |
Sets the nebulizer temperature. Available values: Off, 25° C - 90° C. |
*.Nominal |
Sets the nominal nebulizer temperature. |
*.Value |
Reports the actual nebulizer temperature. |
CarrierFlow |
Sets the carrier gas flow in [slm]. Available values: 0,90 slm - 3,25 slm. |
*.Nominal |
Sets the nominal carrier gas flow. |
*.Value |
Reports the actual carrier gas flow. |
Channel properties and commands:
Command/Property |
Description |
Starts/ Ends data acquisition |
|
Average |
Averages all measured values. |
Reports the signal's slope, i.e., the difference between the value and the value 1 second ago. |
|
Sets the maximum step rate for Step = Auto; default: 5.1 s. |
|
Retention |
Indicates the retention time for the signal in [ms]. |
Signal |
Indicates the signal value in [mV]. |
Sets the step for data acquisition; |
Special commands:
Command/Property |
Description |
AnalogVoltage |
Analog output voltage that corresponds to the current value . |
LightSourceIntensity |
Sets the light intensity of the diode. Possible values: 1% - 100%. |
PMTGain |
Defines an internal gain factor with which the current value is multiplied. However, this value has no effect on the signal-to-noise-ratio. |
SmoothWidth |
Specifies the number of data points used for data smoothing in the detector. |
For a program example, refer to:
Polymer Laboratories ELS 2100/Varian 380-LC ELS Detector: Program Example.