Commands and Tips for Third-Party Devices
Polymer Laboratories ELS 2100 (Ice) / Varian 380-LC (385-LC) ELS Detectors: Commands and Tips

The ELS 2100 and the ELS 380-LC/385-LC drivers support the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):

General commands and tips:

Command/Property

Description

Autozero

Sets the signal output to 10 mV.

AZ_Done

Reports the status of an autozero command (Busy or Done).

ClearDisplayError

Clears any error messages on the detector's display.

Connect

Connects the device to Chromeleon.

Connected

Indicates whether the detector is connected to Chromeleon, that is, under computer control (also, see Connect, Disconnect).

Data_Collection_Rate

Sets the rate at which Chromeleon collects digital data points from the detector (380-LC/385-LC ELS detectors only). Note: This property must be set directly on the device.

Disconnect

Disconnects the detector from Chromeleon and activates manual control from the panel.

DriverID

Unambiguous driver name.

DriverVersion

Indicates the driver version.

FirmwareVersion

Indicates the firmware version.

Manufacturer

Indicates the manufacturer of the driver.

ModelNo

Indicates the instrument's model number.

ModelVariant

Used by Chromeleon to distinguish between the detector variants.

Ready

Ready indicates that the device is ready for operation, i.e., no Autozero is started, the temperature has been reached, no error has occurred.
NotReady
indicates that the device is not ready for operation.

Standby

Sets the detector into the Standby mode or cancels this mode (NoStandby). In standby mode, the detector remains connected to Chromeleon.

 Tip: 

Run the detector in NoStandby mode for successful data acquisition.

 

Device properties:

Command/Property

Description

EvaporatorTemperature

Sets the temperature of the evaporator. Available values: Off, 25° C - 120° C or Off, 10° C - 80° C (ELS 2100 Ice).

*.Nominal

Actual evaporator temperature.

*.Value

Current evaporator temperature.

NebuliserTemperature

Sets the nebulizer temperature. Available values: Off, 25° C - 90° C.

*.Nominal

Sets the nominal nebulizer temperature.

*.Value

Reports the actual nebulizer temperature.

CarrierFlow

Sets the carrier gas flow in [slm]. Available values: 0,90 slm - 3,25 slm.

*.Nominal

Sets the nominal carrier gas flow.

*.Value

Reports the actual carrier gas flow.

 

Channel properties and commands:

Command/Property

Description

AcqOn/Off

Starts/ Ends data acquisition

Average

Averages all measured values.
ON
: Averages all measured values over the step interval (recommended).
OFF
: Records only the last sample of each interval.

Delta

Reports the signal's slope, i.e., the difference between the value and the value 1 second ago.

MaxAutoStep

Sets the maximum step rate for Step = Auto; default: 5.1 s.

Retention

Indicates the retention time for the signal in [ms].

Signal

Indicates the signal value in [mV].

Step

Sets the step for data acquisition;
default: Auto: Selects the best step dynamically (recommended).

 

Special commands:

Command/Property

Description

AnalogVoltage

Analog output voltage that corresponds to the current value .

LightSourceIntensity

Sets the light intensity of the diode. Possible values: 1% - 100%.

PMTGain

Defines an internal gain factor with which the current value is multiplied. However, this value has no effect on the signal-to-noise-ratio.

SmoothWidth

Specifies the number of data points used for data smoothing in the detector.

 

For a program example, refer to:

Polymer Laboratories ELS 2100/Varian 380-LC ELS Detector: Program Example.