Commands and Tips for Third-Party Devices
Perkin Elmer Series 200 Autosampler
Chromeleon provides the commands and parameters from the table below (please note that the display Filter level determines which commands and properties are displayed) to control the Series 200 Autosampler. Also, refer to the information in the autosampler's Operating Instructions.
General Commands and Properties:
Command/Property |
Description |
Connect |
Connects the device to Chromeleon |
Connected |
Indicates whether the instrument is connected to Chromeleon, that is, under computer control (also, see Connect, Disconnect) |
Disconnect |
Disconnects the instrument from Chromeleon and activates manual control from the instrument panel. |
DriverID |
Unambiguous driver name (read-only). |
DriverVersion |
Driver version (read-only). |
Inject |
Performs an injection using the specified parameters (Blank, Position, Volume). |
Manufacturer |
Manufacturer of the driver (read-only). |
Ready |
Indicates whether the autosampler is ready for injection. Values: Ready, NotReady (read-only). |
Status |
Indicates the current status of the autosampler (read-only). |
SyringeVolume |
Indicates the volume of the installed syringe. |
Tray |
Indicates the installed tray type (read-only). |
Wash |
Performs a wash operation. |
Special Commands and Properties:
Command/Property |
Description |
|
AirGap |
Sets the volume [µl] of air drawn before the sample to prevent mixing of the sample and the flush buffer. Range: 1-99 µl. |
|
Derivatization |
Enables or disables derivatization. Values: On (performs derivatization by adding up to 5 different reagents to each sample vial), Off (no derivatization). |
|
DerivMixMode |
Determines the mix mode for derivatization. Values: Standard (no mixing), Intensive (sample is mixed with reagents). |
|
DerivReagentVials |
Sets the number of reagent vials used for derivitazation. Depending on the installed tray type, you can select up to 5 vials. |
|
EndRunWashCycles |
Sets the number of wash cycles after the sample run. |
|
ExcessVolume |
Sets the excess sample volume that is drawn during sample processing. Range: 0-100 µl. |
|
InjectDelay |
Delays sample injection by the selected time to compensate for the dead volume between the pump and the autosampler. Range: 0-999,0 min |
|
InjectMode |
Determines injection mode: Partial Loop (the loop is partially filled with liquid) Full-Loop (the loop is completely filled with liquid) |
|
MixSpeedIn
|
Sets the speed for drawing the mix volume. Values: VerySlow, Slow, Medium, Fast and VeryFast. Select slow or very slow speeds for viscous samples. |
|
MixSpeedOut |
Sets the speed for dispensing the mix volume. |
|
PreInjWashCycles |
Sets the number of wash cycles performed before the sample is drawn and injected. |
|
PostInjWashCycles |
Sets the number of wash cycles performed after the sample was drawn and injected. |
|
SampleHeight |
Determines how far the sample needle moves into the sample vial. The value corresponds to a percentage of the total distance from the shoulder (top) of the sample vial to the bottom of the vial holder.Values: 1% - 99%. 0% is at the bottom of the vial holder, 99% is at the top. |
|
SamplePumpPause |
Sets the time (in seconds) that the autosampler pauses after each sample syringe movement. |
|
Position |
Determines the sample's position on the sample tray. Values: 1-384 (depending on the tray type). Also see Tray. |
|
R1/R2/R3/R4/R5_MixCycles |
Sets the number of cycles for mixing the reagent and sample. Range: 0 - 9. |
|
R1/R2/R3/R4/R5_Position |
Determines the position of the reagent source vial (R1-R5) that is used for derivatization. Values: 1-384 (depending on the tray type, also see Tray). |
|
R1/R2/R3/R4/R5_ReactionTime |
Sets the reaction time of the derivatized sample. |
|
R1/R2/R3/R4/R5_Volume |
Sets the transfer volume for the reagent (R1-R5) used for derivatization. Range: 1-250 µl. |
|
SampleSpeed |
Sets the speed at which the sample syringe aspirates and dispenses sample. |
|
VialVenting |
If vial venting is enabled (On), air is injected into the sample vial before sample is drawn. The air volume is the sample volume plus the buffer volume. Off = no prior injection of air into the vial. |
|
Volume |
Sets the injection volume. The maximum volume depends on the volume of the installed syringe (see table below): |
|
|
Syringe volume: |
Maximum injection volume: |
|
50 µl |
40 µl |
|
100 µl |
83 µl |
|
250 µl |
213 µl |
|
500 µl |
448 µl |
|
1000 µl |
998 µl |
|
2500 µl |
2498 µl |
|
Note: Only integers are allowed as values for the injection volume. |
|
WashCycles |
Determines the number of wash cycles activated by the Wash command. Range: 0 - 9. |
|
WashSpeed |
Sets the speed for washing the needle. Values: VerySlow, Slow, Fast, VeryFast. |
|
WashVolume |
Sets the volume of liquid used for washing the needle before and after each injection. |
Further Information