Commands and Tips for Third-Party Devices
Perkin Elmer TurboMatrix Headspace Sampler: Pressure
Ahe following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
Property |
Range/Values |
Description |
ColumnPressure |
|
Properties related to the column pressure. |
Column Pressure. |
0.00 - 4.00 bar, |
Sets the target pressure in the column. |
Column Pressure. |
0.00 - 4.00 bar |
Indicates the current column pressure (read-only). |
HighPressInject |
On or Off |
Enables or disables high-pressure injection mode (on the device: Hi Psi Inject). Note: This property is available only if you have selected the Programmable Pneumatic (PPC) check box on the Options tab page in the Server Configuration program. |
InjectPressureČ |
0.00 - 4.00 bar |
Sets the pressure for high-pressure injections (HighPressInject = On) for the time between the start of the pressure period and sample venting (withdrawing the needle). Notes: This property is available only if you have selected the Programmable Pneumatic (PPC) check box on the Options tab page in the Server Configuration program. When HighPressInject is set to Off, the device does not use this value. |
Notes:
The sampler accepts Chromeleon pressure commands only in psi. That is why Chromeleon must convert bar or kPa to psig before sending these commands to the sampler. This conversion may result in minor rounding differences, e.g., 120 kPa in Chromeleon is transferred to the sampler as 119.9 kPa.
The range of 0 - 60 psi (or 0 - 413.7 kPa) corresponds to the range that can be set on the device.
For more commands and properties that Chromeleon supports for the TurboMatrix Headspace sampler, refer to: