Commands and Tips for Third-Party Devices
Agilent 6890/6850 GCs: Injector

The following properties are available (please note that the display Filter level determines which properties are displayed):

 

Property

Default

Description

6890 GC

 

 

Mode

² (Demo
mode:
on_column).

Injection mode. Must be set to on_column for cooling on the column inlet. Otherwise, set to Normal.
Usually, the correct setting is automatically selected in the Server Configuration program. It is not necessary to change this setting.

6850 and 6890 GCs

Depth

0.0 mm

Position of the needle in the vial. The default position is about 3.6 mm from the bottom of the vial. The default position is the reference of all other positions.

Fan

² (Demo
mode: Off)

Turns the injector fan on and off.

FirmwareVersion

-

Firmware version (6 characters, read-only)

Inject

 

Injects the sample.

The values Position and Volume can only be used during manual injections. Injection is allowed at  t=0.000 min only.

InjectWaitTime

0.0 s

Time between the inject command and the inject response.

ModelNo

-

Model number (8 characters, read-only)

Position

1

Current sample position on the tablet.

Post-a

0

Indicates how often the syringe is rinsed with solvent from the solvent reservoir A after the injection.

Post-b

0

Indicates how often the syringe is rinsed with solvent from the solvent reservoir B after the injection.

PostDwell

 

Time that the needle remains in the inlet after the injection before it is withdrawn.

Pre-a

0

Indicates how often the syringe is rinsed with solvent from the solvent reservoir A prior to the injection.

Pre-b

0

Indicates how often the syringe is rinsed with solvent from the solvent reservoir B prior to the injection.

PreDwell

0.00 min

Time that the needle remains in the inlet before the piston moves to inject the sample.

Prewash

0

Indicates how often the syringe is rinsed with the sample prior to the injection.

Pumps

0

Indicates the how often the pressure plunger of the syringe is moved up and down while the needle is in the sample. This is to expel gas bubbles and to enhance the reproducibility.

Speed

0

Average piston speed³

State

Off

Indicates whether the autosampler has injected (read-only).

Syringe

0

Indicates the syringe type as specified in the Server Configuration program.

Viscosity

0 s

Number of seconds the pressure plunger of the syringe pauses after the pump and injection strokes. This pause allows viscous samples to flow into the vacuum created in the syringe.

Volume

2 µl

Specifies the amount of sample to be injected (in µl).

The value range provided is from 2% of the smallest possible syringe volume to 50% of the largest possible syringe volume. Also see Syringe.

Waste

² (Demo
mode: Off)

Specifies which waste bottles shall be used (alternating).

 

 Notes:

² When connecting, Chromeleon reads the value from the instrument and accepts it as default.

³ For information about which plunger speed can be achieved with the Speed command for the different syringes, refer to the table below:

 

Syringe (µl)

Plunger Speed (µl/min)

 

Fast

Slow

5

3000

150

10

6000

300

25

15000

750

50

30000

1500

100

60000

3000

250

15000000

750000

500

30000000

1500000

 

In addition, the following commands are available, depending on the filter level:

 

Command

Description

EraseErrorLog

Deletes the sample error log.

StopInject

Stops the running injection.

 

The Administrator Help section provides installation instructions for the 6890 and 6850 GCs; refer to Hardware Installation:

 Agilent 6890 GC: Installation

 Agilent 6850 GC: Installation