Commands and Tips for Third-Party Devices
Agilent 1100/1200 HPLC System: Program Examples for the Fluorescence Detector
The following program example applies to an HPLC 1100 system:
1. One channel, no spectra acquisition
Set the operating mode to ScanMode = Off.
Fluorescence.PeakWidth = 0.200
ScanMode = Off
Emission_1.ExWavelength = 335
Emission_1.EmWavelength = 390
Emission_1.Step = Auto
Emission_1.Average = On
0.000 Wait Sampler.Ready and Fluorescence.Ready
Inject
Emission_1.AcqOn
2.000 Emission_1.AcqOff
End
2. Several channels, no spectra acquisition
Set the operating mode to ScanMode = Emission or Excitation. In addition, set EmScanWhen (and/or ExScanWhen) to None.
Note:
Set ExWavelength (or EmWavelength, respectively) to Off for all channels for which data acquisition will not be performed. The other wavelength must be identical for all channels.
Fluorescence.PeakWidth = 0.200
ScanMode = Emission
EmScanWhen = None
PeakWidthPD = 0.200
Threshold = 1.000
Emission_1.ExWavelength = 335
Emission_1.EmWavelength = 390
Emission_1.Step = Auto
Emission_1.Average = On
Emission_2.ExWavelength = 335
Emission_2.EmWavelength = 450
Emission_3.ExWavelength = Off
Emission_3.EmWavelength = 390
Emission_4.ExWavelength = Off
Emission_4.EmWavelength = 390
0.000 Wait Sampler.Ready and Fluorescence.Ready
Inject
Emission_1.AcqOn
Emission_2.AcqOn
2.000 Emission_1.AcqOff
Emission_2.AcqOff
End
3. Peak-dependent spectra acquisition
Set the operating mode to ScanMode = Emission (or Excitation). In addition, set EmScanWhen (or ExScanWhen, respectively) to Apex (only at the peak maximum) or to AllInPeak (= all spectra during the peaks).
Tip:
Verify that the settings for PeakWidthPD and Threshold match the peaks recorded on the first channel (Emission_1).
Fluorescence.PeakWidth = 0.200
ScanMode = Emission
EmScanWhen = Apex
PeakWidthPD = 0.200
Threshold = 1.000
Emission_1.ExWavelength = 335
Emission_1.EmWavelength = 390
Emission_1.Step = Auto
Emission_1.Average = On
Emission_2.ExWavelength = Off
Emission_2.EmWavelength = 390
Emission_3.ExWavelength = Off
Emission_3.EmWavelength = 390
Emission_4.ExWavelength = Off
Emission_4.EmWavelength = 390
0.000 Wait Sampler.Ready and Fluorescence.Ready
Inject
Emission_1.AcqOn
2.000 Emission_1.AcqOff
End
4. User-defined spectra acquisition
Set the operating mode to ScanMode = Emission (or Excitation). In addition, set EmScanWhen (or ExScanWhen, respectively) to All:
Trigger DoScan Emission_1.Signal > 10.000
Scan
EndTrigger
Fluorescence.PeakWidth = 0.200
ScanMode = Emission
EmScanWhen = All
Emission_1.ExWavelength = 335
Emission_1.EmWavelength = 390
Emission_1.Step = Auto
Emission_1.Average = On
Emission_2.ExWavelength = 340
Emission_2.EmWavelength = 390
Emission_2.Step = Auto
Emission_2.Average = On
Emission_3.ExWavelength = Off
Emission_3.EmWavelength = 390
Emission_4.ExWavelength = Off
Emission_4.EmWavelength = 390
0.000 Wait Sampler.Ready and Fluorescence.Ready
Inject
Emission_1.AcqOn
Emission_2.AcqOn
1.000 Scan
2.000 Emission_1.AcqOff
Emission_2.AcqOff
End
The DoScan Trigger determines that a fluorescence spectrum is recorded when the signal of the Emission_1 channel exceeds 10.
Also, refer to Agilent 1100/1200 HPLC System: Scan Modes for the Fluorescence Detector.
For information about the commands for fluorescence detector control, refer to Agilent 1100/1200 HPLC System: Fluorescence Detector.