Dionex Pumps
ESA Model 582/584 HPLC Pumps
Chromeleon provides the commands and parameters from the table below (please note that the display Filter level determines which commands and properties are displayed).
General commands
Command/Property |
Description |
%A |
Indicates the partial flow of a solvent component, expressed as a percentage of the total flow. |
*Value |
Current value [%] for component A. |
*Equate |
Name of the component (e.g., A). Note: Default component names can be specified on the Properties tab page for the pump in the Server Configuration program. |
Flow |
Sets the flow rate [ml/min]. This is the total flow through the pump. |
*Nominal |
Sets the nominal flow. |
*Value |
Reports the current flow. |
FlowState |
Turns the pump on or off. |
LockState |
Locks or unlocks the keypad of the pump. |
Pressure |
Reports the pressure. |
*Value |
Reports the current pump pressure. |
*.LowerLimit |
Sets the lower pressure limit. The pump will stop the flow if the pressure is below the lower limit. |
*.UpperLimit |
Sets the upper pressure limit. The pump will stop the flow if the pressure exceeds the specified value. |
Pressure channel commands
AcqOn starts data acquisition. AcqOff terminates data acquisition. |
|
Averages all measured values. |
|
Reports the signal slope, that is, the difference between the value and the value 1 second ago. This is useful for Triggers. |
|
Sets the maximum step rate for Step = Auto. |
|
Retention |
Indicates the retention time for the signal. |
Signal |
Reports the current pressure signal. |
Sets the step for data acquisition. |
For additional device commands, refer to General Device Commands.
For installation details, refer to ESA Model 582/584 HPLC Pumps: Installation.