Program Wizard / Program Editor:
Sample Preparation Options: Basic Edit Mode

Details about the Basic Edit Mode available in the AS-DV Sample Preparation Options dialog box are provided below. For an overview of the dialog box, see Sample Preparation Options (AS-DV).

Basic Edit Mode Overview

When you select Basic Edit Mode, a default list of commands is displayed in the Command/Parameters table. The AS-DV executes the commands in the order in which they appear in the table:

You can modify the list of commands by selecting among several options (Volume, Bleed, Rinse). When you select an option, Chromeleon adds the appropriate commands and parameters to the Command/Parameters table. For details about each option, refer to Basic Edit Mode Options below.

Basic Edit  Mode Options

Option

Description

Volume

Specify the volume of sample to deliver from the vial.

  • From Sequence: The AS-DV delivers the sample volume specified in the sequence. You can use the Total Volume Delivered Estimator to calculate the volume of sample taken from the vial.

  • Full: The AS-DV delivers the full vial volume.

  • 1666 ml: The AS-DV delivers 1666 ml from the vial. Use this option if you want to take three injections from a 5000 ml vial.

  • 2500 ml: The AS-DV delivers 2500 ml from the vial. Use this option if you want to take two injections from a 5000 ml vial.

Bleed

Bleed is used to deliver excess sample (and any trapped air) from the vial to waste before delivering the injection volume to the loop or concentrator column. This ensures that only the specified injection volume of sample is delivered to the loop or concentrator column. The bleed occurs only on the first injection from the vial.

To disable the bleed, select None.

To enable the bleed, select the valve that will perform the bleed and the valve position that will direct the bleed to waste.

  • To use the injection valve, select either LoadPosition or InjectPosition.

  • To use the AS-DV auxiliary valve, select AuxValve.A or AuxValve.B.

 Tip:

If the valve is plumbed according to the standard plumbing described in the AS-DV operator's manual, select InjectPosition (if you are using an injection valve) or AuxValve.A (if you are using an auxiliary valve).

The selected valve command and a DeliverSample Volume = Bleed command are added to the Command/Parameters table.

Rinse
 

To run a rinse cycle before the injection, select the Rinse check box option.

Rinse Volume

Enter the desired Rinse Volume. The volume is taken from the vials in the Rinse Vial Range, specified in the Server Configuration program. To define a rinse vial range, see Options (AS-DV Autosampler Dialog Box).

An InjectPosition command and a DeliverRinse command are added to the beginning of the Command/Parameters table.