Program Wizard / Program Editor:
Sample Preparation Options: Advanced Edit Mode
Details about the Advanced Edit Mode available in the AS-DV Sample Preparation Options dialog box are provided below. For an overview of the dialog box, see Sample Preparation Options (AS-DV).
Advanced Edit Mode Overview
When you select Advanced mode, the Command/Parameters table becomes active. You can add, delete, and modify commands and select additional commands and parameters not available in Basic Edit Mode.
Notes:
The EndSamplePrep and Inject commands cannot be deleted.
The Inject command is always the last command in the list.
Advanced Edit Mode Options
Option |
Description |
Command |
Lists the available commands: LoadPosition: Sets the injection valve to the load position. InjectPostion: Sets the injection valve to the inject position. DeliverRinse: Delivers rinse fluid through the sample out line. Enter the Volume of rinse to deliver and the Vial from which to take the rinse volume. DeliverSample: Delivers sample through the sample out line. Enter the Volume of sample to deliver and the Vial from which to take the sample. The following two commands are only available if the optional valve is installed and configured as an auxiliary valve. AuxValve.A: Sets the valve to the A position. AuxValve.B: Sets the valve to the B position. |
Volume |
For DeliverRinse:
For DeliverSample:
|
Vial |
For DeliverRinse:
For DeliverSample:
|
Insert |
Inserts a new command before the command currently selected in the Command/Parameters table. |
Append |
Adds the command and parameters selected in the Command list to the Command/Parameters table before the Inject command. |
Modify |
Replaces the command currently selected in the Command/ Parameters table with the command and parameters selected in the edit fields above the table. |
Delete |
Removes the command currently selected in the Command/ Parameters table. |