Installing and Controlling Third-Party Devices
Waters Alliance 2690(D)/2695(D) HPLC Modules: Installation

 

Device Type:

HPLC module

Device Driver:

Waters: 2690/2695 Separation Module

The driver was tested with firmware version 1.22.

Waters: 2690D/2695D Separation Module

The driver was tested with Alliance firmware version 2.00 and Hanson SR8Plus software version 3.10

Hardware Options
Supported:

2690(D)/2695(D) or 2690/2695 XE (firmware 1.20 or higher):

  • Pump: Quaternary gradient pump (0.05 to 5.00 mL/min)

  • Injector (120 vials, 5 carousels)

  • Sample Temperature Control (XE version: 4-40°C)

  • Column heater

    2690(D) with column heater (20-60°C)

    2695(D) with column heater (20-65°C)

    2695(D) with column heater/cooler (4-65°C)

     Tip:  

    The column heater type is not read from the instrument automatically and must be configured manually by the user.

  • In-Line Vacuum-Degasser with relative pressure reducer (XE version) or He degasser

2690D/2695D Separation Module:

  • Separation module with 6 or 8 needle dissolution dispenser

  • Transfer module with 6 or 8 syringes

  • Up to 2 Hanson SR8-Plus baths or compatible baths of other manufacturers

Alliance eSeries:

The Alliance e2695 Separation Module is equipped with a GPIB communication interface which allows control of the system using the 2690(D)/2695(D) driver.

Restrictions:

  • All parameters related to the flow, temperature, and event control are downloaded to the instrument as one single block. As a result, most properties cannot be changed at negative retention times.

  • Thus, the Waters Alliance 2690(D)/2695(D) HPLC modules do not support sample preparation.

  • The Waters Alliance 2690(D)/2695(D) HPLC modules must first be completely initialized before they are ready to use.

  • If you condition the column via the associated dialog on the device, it may happen that the PGM file cannot be executed.

  • For firmware versions < 2.00, the syringe speed settings Slow and Normal may be interchanged. Therefore, Dionex recommends upgrading the separation module to firmware version 2.00 or higher.

  • This Waters device cannot connect to a NI-488.2 driver version 15.0 or later. You must use a NI-488.2 driver with software version ≤ v.14.0.

  • Per GPIB card (GPIB bus) only one Waters Alliance HPLC system is allowed.

  • Due to GPIB data rate limitations, only one Waters 996 PDA detector per Alliance system and GPIB bus is allowed.

2690D/2695D Separation Module only

  • Only one 269xD Alliance system is supported per timebase

  • Only one dissolution test per sequence can be run (applies to single and dual-bath systems)

  • Dissolution test and LC analysis must be performed within the same sequence

  • An abort of a dissolution test on the Alliance system is not reported to Chromeleon

 

What is required?

License:

Timebase Class 2

Connection:

The following cable is required:

GPIB cable (Thermo Scientific/Dionex part no.: 5919.0013; for more information refer to  GPIB IEEE 488 Bus Cable)

 Tip: 

You cannot connect the Waters Alliance 2690(D)/2695(D) HPLC Modules to a UCI Universal Chromatography Interface because the UCI does not provide a GPIB connector.

Hardware
Prerequisites:

GPIB interface card or GPIB/USB adapter.

 Tips:  

The Waters BusLAC/ETM card cannot be used with Chromeleon.

Dionex advises against using Waters devices and Agilent HP 1100 Systems on the same GPIB card.

Software
Prerequisites:

VISA library (≤ version 14.0)

Control Panel:

Waters2690_996PDA.pan

Hardware Installation

  1. First, install Chromeleon software, and then connect the instrument using the GPIB IEEE 488 bus cable from Dionex.

  2. Make the following settings on the Alliance module:

For further settings refer to the general descriptions of commands under Commands and Tips for Third-Party Devices  Waters Alliance 2690(D)/2695(D) HPLC Modules.

 Notes:

Software Installation

For details about how to add timebases and device drivers to the Server Configuration, refer to Adding, Configuring or Deleting Devices. When you add a driver to the timebase, the configuration dialog or wizard opens automatically. For information about the settings, click the Help button on a page or press F1.

Troubleshooting

 Waters Instruments: Troubleshooting

 GPIB-Connected Devices: Troubleshooting

For a list of all error messages, refer to the Error Levels tab page of the instrument's Properties in the Chromeleon Server Configuration program.

Further Information

For more installation details, please refer to the manual for the instrument.

For information about the supported GPIB interfaces, refer to Connecting Third-Party Devices via GPIB.

For an overview of the different Waters instruments that can be controlled by Chromeleon, refer to  Waters.