Installing and Controlling Third-Party Devices
Shodex HPLC Refractive Index Detectors (RI-101, RI-102, and RI-104)
Device Type: |
Single-channel refractive index detector |
Device Driver: |
Shodex: RI-101/102/104 RI Detector |
Hardware Options |
RI-101: Analytical version Tip: The Shodex RI-101 Refractive Index Detector is available from Dionex as part of the Summit HPLC series. RI-102: Preparative version RI-104: Semi-micro version |
Restrictions: |
With digital data acquisition, the maximum duration of an analysis (i.e., of a single sample) is limited to about 1,500 minutes (~25 hours). Runs which exceed this period are terminated with a communication timeout error. For runs that take longer than 25 hours it is recommended to use analog data acquisition, for example, via an UCI-100 interface. |
What is required?
License: |
Timebase Class 1 |
Connection: |
A free RS232 port and the following items are required: |
Firmware Version: |
Recommended firmware version: 1.10 or later. Note: With earlier firmware versions (1.09 and earlier), communication errors may occur if the refractive indexes exceed the maximum RI range (OVER) (indicated by the error message "A communication timeout occurred while waiting for device answer in TaskGetAquData"). A problem that occurs with older devices is that the correct firmware version cannot be communicated to the software. As a result, the device always displays 1.06 as firmware version even if a more current version is installed. This problem also occurs if the following chip versions are installed:
With these chip versions installed, the actual firmware version is 1.09. However, the detector displays firmware version 1.06 on the startup screen and communicates this version to Chromeleon. |
To control the instrument, use the Dionex_LC/Shodex_RI101.pan from the Dionex Templates\Panels\LC_Other folder in the Browser. |
Hardware Installation
Device Connection: |
First, install Chromeleon software, and then connect the instrument to the RS232 port on the Chromeleon server PC. |
Software Installation
For details about how to add timebases and device drivers to the Server Configuration, refer to Adding, Configuring or Deleting Devices. When you add a driver to the timebase, the configuration dialog or wizard opens automatically. For information about the settings, click the Help button on a page or press F1.
Further Information
For detailed installation instructions, refer to the operating instructions for the instrument.
For troubleshooting information, refer to Shodex Refractive Index Detectors (RI-101, RI-102, RI-104).