Installing and Controlling Third-Party Devices
Agilent 1200 HPLC System: Supported Hardware Options
The table provides an overview of the supported hardware options and required firmware versions.
Notes:
The firmware versions listed in the table are the minimum requirements. Dionex recommends always using the latest firmware version. Chromeleon has been tested with firmware B.06.10 and A.06.10.
The Agilent 1200 driver is not compatible with firmware version A.05.xx and earlier.
Instrument |
Module No. |
Firmware |
Comment |
Detectors: |
|
|
|
Variable Wavelength |
G1314A G1314B G1314C (SL type) G1314D G1314E |
A.06.10 A.06.10 A.06.10 B.06.21 B.06.21 |
Not tested by Dionex.
|
LC Diode Array Detector (DAD) |
G1315A G1315B G1315C (SL type) G1315D |
A.06.10 A.06.10 B.06.10 B.06.10 |
Die "B" Firmware-Versionen sind nicht mit den in Chromeleon verfügbaren Steuerfenstern zur Durchführung des Lampenintensitätstest kompatibel. |
Fluorescence Detector (FLD) |
G1321A |
A.06.10 |
|
Refractive Index Detector (RID) |
G1362A |
A.06.10 |
Not tested by Dionex. |
Multiple Wavelength Detector (MWD) |
G1365A G1365B |
A.06.10 |
Model variants G1365A/B/C have not been tested by Dionex. Some of the functions and commands described for the 1200 HPLC system may therefore not be supported or are subject to certain limitations. |
|
G1365C (SL type) |
B.06.10 |
|
|
G1365D |
B.06.10 |
|
|
|
|
|
Pumps + Degassers: |
|
|
|
Isocratic Pump |
G1310A |
A.06.10 |
Not tested by Dionex. |
Low-pressure gradient pump |
G1311A |
A.06.10 |
|
High-pressure gradient pump with (optional) solvent selection valves |
G1312A G1312B (SL type) |
A.06.10 |
|
Degasser |
G1322A G1322B |
n.a. |
|
Degasser |
G1379A G1379B |
n.a. |
|
|
|
|
|
Autosamplers: |
|
|
|
Autosampler |
G1329A |
A.06.10 |
|
Injection Unit |
G1327A thermostatted unit (= G1329A autosampler + G1330A thermostat) |
A.06.10 |
|
Wellplate LC Autosampler |
G1367A G1367C + G1330B (thermostat) G1367D (SL type) |
A.06.10 |
|
Sample thermostat |
G1330A G1330B |
A.06.10
|
|
|
|
|
|
Thermostated Column Compartments: |
|
|
|
Column compartment with (optional) column thermostat + column switching valve |
G1316A G1316B (SL type) G1316C (SL type) |
A.06.10 |
|
|
|
|
|
Fraction Collector |
G1364B G1364C |
A.06.10 |
Rinsing and septum piercing not supported by the G 1364B. Tip: To avoid known firmware issues, the following firmware versions should not be used: A.06.32, A.06.33. As an alternative, you may use firmware version A.06.10 or A.06.34. |
|
|
|
|
Controller: |
|
|
|
Instant Pilot |
G4208A |
B.02.05 |
|
|
|
|
|
External Valves: |
|
|
Valves marked with a (*) are not yet tested by Dionex |
6-position/7-port valve |
G1156A* |
A.06.10 |
|
2-position/10-port valve |
G1157A* |
A.06.10 |
|
2-position/6-port valve |
G1158A |
A.06.10 |
|
2-position/6-port valve, high pres. |
G1158B* |
A.06.10 |
|
6-position selection |
G1159A |
A.06.10 |
|
12-position/13-port valve |
G1160A* |
A.06.10 |
|
2-position/6-port micro valve |
G1162A* |
A.06.10 |
|
2-position/10-port micro valve |
G1163A* |
A.06.10 |
|
Firmware Compatibility
Beginning with revisions A.06.1x and B.06.1x, the following rules apply:
All minor revisions within a decade (e.g., A.06.10 to A.06.19 or B.06.10 to B.06.19) are compatible among themselves and may be mixed within a HPLC stack. The same applies to revisions of decades A.06.2x / B.06.2x, and A.06.3x / B.06.3x.
A and B revisions are compatible if they are of the same decade, for example, A.06.10 and B.06.12.
All revisions of different decades, for example, A.06.1x versus A.06.2x, or B.06.1x versus B.06.2x, are not compatible with each other and must not be used within the same HPLC stack.
Further Information
For installation details, refer to the instrument manual and to:
1200 HPLC System: Installation.
For information about the requirements when installing the 1200 HPLC System via a LAN connection, refer to:
Agilent 1100/1200 HPLC System: Racks for information about the available rack types.
For an overview of the different Agilent/HP instruments for which device drivers are available in Chromeleon, refer to Agilent (formerly HP).