Installing and Controlling Third-Party
Devices
Applied Biosystems ABI 785A UV Detector: Installation
Device Type: |
UV Detector |
Device Driver: |
ABI: ABI_785A_UV_VIS_Detector |
Supported Hardware Options: |
No special options available. |
What is required?
License: |
Timebase Class 2 |
Connection: |
RS-232
cable. (For more information, refer to |
Hardware |
|
Abi785A.pan |
Hardware Installation
Software Installation
Install the Device Driver in the
Chromeleon Server Configuration. (For more information about the program,
refer to Software Installation and Communication
The
Server Configuration
Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the detector to the timebase. Select Add Device on the Edit or context menu, and then select ABI from the Manufacturers list box and ABI_785A_UV_VIS_Detector from the Devices list box.
The following settings are required:
On the ABI_785A_UV_VIS_Detector tab page, verify that the Demo Mode is disabled.
On the Generic tab page, select a free COM port.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the ABI 785A UV Detector, connect the Abi785A.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
Rise Time, Scan, and SetRelays are available as special parameters (however, Scan and SetRelays are not included in the ABI785A.PAN control panel).
Further Information
Tip:
For more installation details, refer to the manual for the instrument.
For an overview of manufacturers whose instruments
can be controlled by Chromeleon, refer to Installing
and Controlling Third-Party Devices.