%A, %B, %C, %D_Level (Solvent Volume)
Instrument Type: |
Pump |
Related Commands: |
|
Description: |
This command monitors solvent consumption. |
Function: |
Before starting a sequence, enter the existing volume of each solvent component. Based on the flow rate and the solvent composition, Chromeleon calculates whether the amount of available solvent is sufficient for the sequence. |
Parameters: |
|
Value |
Available volume of the corresponding component in [l]. |
LowerLimit |
Lower limit for the corresponding component in [l]. If this value is reached, the batch is aborted and the following message appears: [Abort] xx:xx:xx {Pump} The %A level is below the limit (x l)! Please refill %A. Tips: In the emergency program, reduce the flow rate to 0 ml/min. This ensures that the pump and the detector cell never run dry. |
Tip:
|
Chromeleon checks the expected demand for solvent during the Ready Check. If not enough solvent is available, a warning is displayed. |
Note:
This command is not available for all ion chromatography pumps.