GC Control
Injecting Two GC Samples Simultaneously (TRACE 1300 Series GC)
To double the sample throughput, the TRACE 1300 Series GC supports the dual-operation mode, which lets you simultaneously inject two samples from different positions.
General Guidelines for Dual Operation Methods
The following conditions must be met:
The GC must be fitted with the required hardware components.
The GC must be installed in two timebases (refer to Thermo Scientific TRACE 1300 Series GC: Dual Operation).
Seperate PGM files must be created for the "Front" and "Back" timebases.
The PGM runtime of both PGM files must be the same.
The oven temperature ramp must be identical.
All auxiliary heaters can be shared, depending on the location property defined in the configuration.
After installing the GC, follow the steps below to set up the dual-operation mode.
Creating two PGM Files
Create a PGM File for each timebase, observing the following:
Enable dual operation with the following command (this can be done only from the control panel):
|
DualOperation |
= Shared |
Enable dual operation for the related sample(s):
|
Sharable |
= ON |
Make sure that the following properties shared by the timebases have the same values at the same retention times: If the values and/or retention times are different, the samples will be processed one after the other.
- PostRunTempCtrl
- PostRunTemperature
- PostRunTime
- CryoCtrl
- StartCryo
- CryoTimeout
- PrepRunTimeout
- AutoStart (service level)
- EquilibrationTime
- ReadyDelay
- AuxHeater.TempCtrl (if aux heater is shared)
- AuxHeater.Temperature (if aux heater is shared)
- TempCtrl
- Temperature.Nominal
Notes:
The sample types are checked by Chromeleon. It is not possible to run a blank sample and a sample of another type in parallel. If one sample is a blank, the other must be as well.
Sharing is not possible for emergency PGM files. Emergency programs are not checked by Chromeleon but in case of an emergency, two programs are sent to the driver, which will then be sent to the GC. The program creator is responsible for providing consistent emergency programs.
Creating a sequence and a batch for each timebase
The batches do not have to include the same number of samples. If all samples of one batch are processed, sample processing will continue for the remaining samples of the other batch.
Tip:
You can add new samples or sequences to the finished batch and restart the batch. Sample processing is resumed when the next sample of the running batch is started.
Starting batches
In dual operation mode, a batch must be started on both timebases simultaneously:
Create the two batches and perform the Ready Check on both timebases.
Start the batch on the first timebase. A message may appear in the Audit Trail telling you to start a batch on the other timebase, too
Start the batch on the other timebase.
The two batches run in synchronization. The Ready Check checks that the shared properties, temperature gradient, and run time for the sample pairs match and that the injections are performed from different positions.
Note:
Observe the following restrictions:
The Ready Check can check only the active sequences.
The Ready Check can detect changes to the running sequence on the other timebase only when you stop and restart the batch on the other timebase. If the Ready Check reports an error on the other timebase, follow the steps below.
- In the timebase for which the error was reported:
- Stop the batch.
- Eliminate the error.
- For the first sample, change the status from Interrupted to Single, if necessary.
- Restart the batch.
- Return to the original timebase.
- Run the Ready Check again.
Examples
a) To analyze a pair of sequences (e.g., Front1 and Back1) simultaneously on two timebases (e.g., Front and Back) and then run another sequence (Front2) only on the Front timebase:
For the pair of sequences, set Sharable = On.
In the PGM File for the Front2 sequence, set Sharable = Off.
b) Use the following settings if you want to process samples in dual operation mode whenever possible but also analyze them one after the other if required:
Accept the default setting: DualOperation = Shared, however
Do not enter the Sharable command in the PGM File.
c) If you want to run a batch only on one timebase (e.g., on the Front timebase), select one of the following alternatives:
In the PGM File, set Sharable = Off and start the batch on the Front timebase.
--Or--
On the control panel, set DualOperation = FrontOnly and start the batch on the Front timebase.
In both cases, the batch is started immediately and does not wait for a batch on the Back timebase.
For information about how to install the TRACE 1300 GC Series, refer to Thermo ScientificTRACE 1300 Series GC: Installation.