Agilent LC/GC System Control (ICF)
Miscellaneous Hints

This topic describes practical hints and known issues with Agilent LC/GC systems controlled by the Instrument Control Framework (ICF).

 

 

  1. Fraction collection is not supported for the Agilent ICF.

 

The following scenarios apply to the Agilent 7697A Headspace Sampler only:

  1. Some instrument parameters logged to the audit trail are rounded to the nearest integer value. However, all values are downloaded to the instrument with the proper precision.

  2. When instrument method parameters of the 7697A headspace sampler are included in a report, the Fill Pressure parameter is rounded to the nearest integer value.

  1. If you use the front panel of the Agilent 7697A headspace sampler, the allowed range for the diameter of the Transfer line parameter is 200-600 microns. However, in the Chromeleon device configuration for the sampler the limit of this value is 250-530 microns.

  1. If you close the PGM Editor and reopen it to adapt your settings, such as the purge flow or time, the Save button is not enabled and your changes will not be saved after closing the editor. This is a limitation of the Agilent ICF Framework for GCs. To avoid losing your changes, select a different tab page and click the Save button.

 

Further Information

Agilent LC System Commands (ICF)

Agilent GC System Commands (ICF)

Creating, Modifying, Resolving and Printing Methods

Setting the Run Time

Monitoring the Instrument Status

Troubleshooting