Commands and Tips for Third-Party Devices
Waters 600E Pump
In addition to General Device Commands, the commands and properties from the table below are available for the Waters 600E pump (please note that the display Filter level determines which commands and properties are displayed).
Also, refer to the information in the operating instructions shipped with the instrument.
General Properties:
Property |
Description |
PumpHeadVolume |
Indicates the volume of the configured pump head (100 µl or 225 µl). |
Specific Commands and Properties:
Property |
Description |
Percentage of each solvent component. |
|
*.Equate |
Name of the component (A, B, C, or D). |
*.Value |
Current value [%] for components A, B, C, or D (read-only). |
DegasFlow |
Specifies the sparge flow to degas the reservoirs. Available range: 0–100 ml/min. |
DegasReservoirA/B/C/D |
Turns the sparge flow for reservoirs A–D ON or OFF. |
Sets the flow rate [ml/min]. Range: 0–20 ml/min (100 µl pump head) or 0–45 ml/min (225 µl pump head) |
|
*.Nominal |
Reports the nominal flow. |
*.Value |
Reports the current flow (read-only). |
Pressure |
Indicates the pump pressure. Range: 100 µl pump head: 0–6000 psi (420 bar) or 225 µl pump head: 0–1000 psi (70 bar) |
*.LowerLimit |
Sets the lower pressure limit. The system stops the sample batch and starts emergency handling if the value is outside the set range. Range: 100 µl pump head: 0–5950 psi (417 bar) or 225 µl pump head: 0–950 psi (67 bar) |
*.UpperLimit |
Sets the upper pressure limit. The system stops the sample batch and starts emergency handling if the value is outside the set range. Range: 100 µl pump head: 51–6000 psi (3.5–420 bar) or 225 µl pump head: 51–1000 psi (3.5–70 bar) |
*.Value |
Current pump pressure (read-only). |
For information on creating a PGM for the pump, refer to Waters 600E Pump: Program Example.