Commands and Tips for Third-Party Devices
Waters Alliance 2790/2795: Pump and Degasser
In addition to the standard commands (refer to Practical Tips for Device Control Pump and Flow Control), the Waters 2790/2795 Separation Module device driver supports the following pump commands and properties (please note that the display Filter level determines which commands and properties are displayed):
Pump Commands and Properties
Property |
Min |
Max |
Description |
BubbleDetection˛, |
Off |
On |
Checks the system for gas bubbles in the flow path, can return an error message and perform a bubble reduction procedure. The following settings are required on the instrument: In the Default method, set the Alarms/Bubble detect field to Alert User, Stop Funct, or StopFlow as described in the instrument's manual on page 1 (Mobile Phase). |
Continue |
|
|
see StopFlow |
Curve |
convex |
concave |
Gradient Curve (Default: linear) Note that the figures used to describe the gradient curve shape in Chromeleon differ from the figures used in the Waters Alliance software. The table below shows how the Chromeleon (CM) curve figures 1–9 are mapped to the Waters Alliance figures 1–11: Curve no. CM –> Curve no. Alliance: n/a –> 1 1 –> 2 (convex) 2 –> 3 (convex) 3 –> 4 (convex) 4 –> 5 (convex) 5 –> 6 (linear, default) 6 –> 7 (concave) 7 –> 8 (concave) 8 –> 9 (concave) 9 –> 10 (concave) n/a –> 11 (No equivalences for Alliance figures 1 and 11 in Chromeleon). Tip: If the gradient defined by Curve is steeper than the one defined by the MaxAcceleration parameter, the actually performed gradient can be different from the gradient defined here. In this case, a warning appears. This is always the case for step gradients. |
MaxAcceleration˛ |
0.01 min |
30.00 min |
Sets the time needed by the pump to reach the maximum flow rate of 10 ml/min. The property serves to protect the column from major pressure changes. It is connected with the MaximumFlowRamp property: MaxAcceleration = (Default: 0.01 min. The property is visible only on the Advanced level.) |
MaximumFlowRamp |
0.34 |
1000.00 |
Maximum flow rate increase in [ml/min˛]. This property serves to protect the column from major pressure changes. This property is connected with the MaxAcceleration property as follows: MaximumFlowRamp = (Default: 1000.00 ml/min˛. The property is visible only on the Advanced level.) |
0.0 bar |
344.70 bar |
Upper and lower pressure limits. For the pressure limits to take effect, the following setting is required on the instrument: In the Default method, set the Alarms/Min. and Alarms/Max fields to Alert User, Stop Funct, or StopFlow as described in the instrument's manual on page 1 (Mobile Phase). The limits entered last in Chromeleon are valid (not the values set on the instrument). Note: The instrument checks the lower pressure limit only during injection. |
|
Pressure.Value |
0.0 bar |
380 bar |
Indicates the current pump pressure. |
PrimaryPressure |
-3.45 bar |
380 bar |
Pressure after the first piston chamber (The property is read-only and visible only on the Expert level.) |
StrokeVolume˛ |
Auto; 25 µl; |
Volume delivered with each piston stroke (Default: Auto. The property is visible only on the Advanced level. ) |
|
|
|
When using these commands for the pump of the Alliance system, please note: When the Stop command is executed while a gradient program is running, the Continue command does not restart the gradient. Instead, the pump is started with constant solvent composition and constant flow rate. |
Tip:
˛ It is not possible to change the property during an injection, during gradient runs, or at negative retention times.
Note:
As described in the instrument's manual, negative pressure readings are possible for the pressure behind the first piston chamber (PrimaryPressure property).
Degasser Commands and Properties
The standard degasser of the Waters Alliance 2790/2795 HPLC modules is a Helium degasser. Select the following property to specify which solvent reservoirs shall be helium sparged and set the sparge rate for each reservoir:
Property |
Min |
Max |
Default |
Description |
HeliumSparge_A |
0 |
100 |
0 |
If the sparge rate is set to 100% for a specific reservoir, the solvent in this reservoir is helium sparged all the time. If the sparge rate is less than 100%, the corresponding valve turns on and off according to the specified percentage. |
The standard degasser of the Waters Alliance 2790/2795 XE HPLC modules is a vacuum degasser (In-Line Degasser). Specify the degassing mode:
Property |
Value |
Default |
Description |
Degasser |
Normal |
Off |
Normal: The degasser turns on and off, depending on the pressure. |
Tip:
˛ It is not possible to change the setting during an injection, during gradient runs, or at negative retention times.
The Pressure property indicates the current pressure:
Property |
Min |
Max |
Description |
Degasser |
-1.0 |
1.0 |
Indicates the pressure (in bar) measured by the degasser. (The property is read-only.) |
For information about how to install the Waters Alliance 2790 and 2795 HPLC modules, refer to Hardware Installation Waters Alliance 2790/2795 HPLC Modules: Installation in the Administrator Help section.