Commands and Tips for Thermo Scientific Devices
Finnigan/TSP UV2000 Detector

The following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):

 

Property

Min

Max

Default

Description

DetectionMode

-

-

-

Sets the wavelength mode:

Single Mode: 190 - 800 nm
(only one detector channel is used)

Dual1 Mode: 190 ... 450 nm

Dual2 Mode: 366 ... 800 nm

Response

0.0

5.0

1.0

Time constant. Indicates the time needed by the detector to reach 63% of the full-scale deflection.

 

Property

Description

UV_LampOn

Turns on the deuterium lamp.

UV_LampOff

Turns off the deuterium lamp.

Visible_LampOn

Turns on the tungsten lamp.

Visible_LampOff

Turns off the tungsten lamp.

 

Channel-Specific Commands and Properties

 

Property

Min

Max

Default

Description

Range

0.0005 AUFS

3.0000 AUFS

1.0000 AUFS

Amplifier gain for the output signal

Wavelength

190 nm

800 nm

254 nm

Wavelength at which the detector measures the signal. Legal range: see DetectionMode.