Commands and Tips for Thermo Scientific Devices
ThermoQuest AS3000 and AI3000 Autosampler Commands
Chromeleon supports the following special commands and parameters for these autosamplers (please note that the display Filter level determines which commands and properties are displayed). Also, refer to the information in the manual supplied with the instrument.
Command |
Description |
Clean |
Cleans the syringe using the previously defined parameters (see below). |
Inject |
Injects the sample using the previously defined parameters (see below). Tip: For these autosamplers, 5µl and 10µl syringes are supported by Chromeleon (also refer to chapter ThermoQuest AS3000 and AI3000 Autosamplers: Installation). According to technical specification, the maximum injection volume must not exceed 50% of the syringe capacity. |
Recalibrate |
Performs the start procedure again, allowing the autosampler to recalibrate the internal positions, such as injector, wash vials, tray. This command is required when a new tray was installed. |
StopInject |
Aborts a running injection. |
Use the following parameters to specify the Inject command in detail. Define all injection parameters before initiating the Inject command. Otherwise, they will not take effect!
Parameter |
Description |
AirVolume |
Select this parameter to specify the volume of air (0.0 - 3.0 µl) drawn after pulling the syringe needle out of the vial. This reduces evaporation from the syringe needle. The total of AirVolume plus Volume must not exceed the syringe volume. |
DrawSpeed |
Select this parameter to determine how fast the sample is drawn (slow or fast). |
FillStrokes |
Select this parameter to determine how often the syringe plunger is moved up and down before the sample is actually drawn up (0...15). This eliminates bubbles and thus, enhances reproducibility. |
InjectWaitTime |
Set the time between the injection command and the response from the autosampler (0 - 10.000.000 s) |
PostInjDelayTime |
Use this parameter to determine for how long the syringe remains in the injector after the injection (0 - 63 s). Note: You can also use the PostInjDelayTime and PreInjDelayTime properties to specify the needle depth during injection. The following settings are available: Standard: PostInjDelayTime or PreInjDelayTime > 0. Minimum: PostInjDelayTime and PreInjDelayTime = 0. |
PostWash |
Select this parameter to set the number of post-injection cleaning cycles (0 - 15). |
PostWashVial |
Specify which vial(s) is(are) used for washing the syringe after the injection with the solvent(s): A, B, C, D, A+B, C+D. |
PreInjDelayTime |
Use this parameter to determine the delay the syringe remains in the injector before the injection (0 - 63 s). Note: You can also use the PreInjDelayTime and PostInjDelayTime properties to specify the needle depth during injection. The following settings are available: Standard: PostInjDelayTime or PreInjDelayTime > 0. Minimum: PostInjDelayTime and PreInjDelayTime = 0. |
PreWash |
Select this parameter to set the number of pre-injection cleaning cycles (0 - 15) |
PreWashVial |
Specify which vial(s) is(are) used for washing the syringe before the injection with the solvents: A, B, C, D, A+B, C+D. |
SampleDepth |
Specify how deep the needle descends into the vial when drawing the sample for injection (bottom or half). |
SampleWash |
Select this parameter to determine the number of cleaning cycles with sample before injection (0...15). |
State |
Indicates the status of the autosampler (read-only): |
For more information, refer to Hardware Installation ThermoQuest AS3000 and AI3000 GC Autosamplers.