Commands and Tips for Third-Party Devices
Shimadzu GC-2010/2014 GC System: General
Please note that the display Filter level determines which commands and properties are displayed:
Command/Property |
Range/Values |
Description |
AmbientTemperature |
-5°C to 80°C |
Reports the ambient temperature (read-only). |
AtmosphericPressure |
47 to 107 kPa |
Reports the atmospheric pressure (read-only). |
Connect |
|
Connects the device to the data system and enables control from the PC. The connect command fails if a method is started or another software controls the GC. |
Connected |
Connected/Disconnected |
Indicates whether the device is connected to Chromeleon, i.e., whether it is controlled from the PC (see Connect, Disconnect). |
Disconnect |
|
Disconnects the device from the data system and enables manual control from the device's control panel. |
EquilibrationTime |
0 min to 9999.99 min |
Indicates the time that the GC waits after reaching the target temperature. |
ErrorCode |
0 to 1000 |
Indicates the error code if an error occurred (read-only). |
Event91 |
Set, NotSet |
Switches the related relay. |
Event92 |
Set, NotSet |
Switches the related relay. |
FirmwareVersion |
|
Indicates the firmware version of the device (read-only). |
GC-System |
On/Off |
Activates the GC according to the parameters on the system's main screen. |
OverTemp |
OK, Fault |
Indicates the temperature status of the column oven (read-only). |
Ready |
Ready/NotReady |
Indicates whether the GC is ready for injection (read-only). Enter a Wait GC Ready command directly before the injection. This ensures that the temperature gradient is performed in due time. |
Reset |
|
Restores the default settings. |
RunState |
Ready, NotReady, Diagnosis, Baking, Run, Pre-Run, Post-Run |
Reports the GC status (read-only). |
SerialNo |
|
Reports the serial number of the GC (read-only). |
TempCtrl |
On/Off |
Turns column oven's temperature control on or off. |
Temperature |
|
Sets the oven temperature [°C]. |
Temperature.Nominal |
0°C to 450°C |
Sets the target operating temperature for the column oven. The temperature ramp can be programmed with a maximum rate of up to 250°C/min and up to 20 steps (not counting the first one). |
Temperature. |
50°C to 450°C |
Determines the maximum allowed oven temperature. The GC interrupts the power supply for the oven if the temperature exceeds the upper limit. |
Temperature.Value |
0°C to 450°C |
Reports the actual oven temperature (read-only). |