Commands and Tips for Third-Party Devices
Shimadzu GC-2010/2014 GC System: Detectors

In addition to the standard detector commands (see  Dionex Detectors), the following commands and properties are available for the various detector types (please note that the display Filter level determines which commands and properties are displayed):

Commands and properties for all detectors:

GC-2014: Note that the settings for the flow rate of make-up gas are only available if the advanced Pressure Control (APC) is installed.

 

Command/Property

Range/Values

Description

AutoAutoZero

On, Off

If activated, autozero is performed automatically.

DataCollectionRate

1, 2, 5, 10, 25, 50 Hz

Shows the data collection rate that was selected in the configuration (read-only).

DetectorCtrl

On, Off

Turns the detector on or off.

DetectorType

Unknown, FID, ECD, DFID, DTCD

Indicates the selected detector type (read-only).

FilterTimeConstant

4 ms, 5 ms, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s

Sets the time constant for the noise filter of the detector.

MakeUpCtrl

On, Off

Turns the makeup gas flow control on or off.

MakeUpFlow

0.0 to 1200.0 ml/min

Sets the makeup gas flow.

MakeUpGasType

He, N2, H2, Ar

Enter the current gas type of the makeup gas.

Temperature

 

Sets the detector oven temperature.

*.Nominal

depends on detector type

Sets the nominal detector oven temperature.

*.UpperLimit

depends on detector type

Sets the maximum detector oven temperature.

*.Value

depends on detector type

Indicates the actual detector oven temperature (read-only).

 

Special commands and properties for the (Dual) Flame Ionisation Detectors (FID, DFID):

GC-2014: Note that the settings for  the flow rate of hydrogen and air are only available if the advanced Pressure Control (APC) is installed.

 

Command/Property

Range/Values

Description

AirCtrl

On, Off

Turns the air flow control on or off.

AirFlow

0.0 to 1200.0 ml/min

Sets the air flow.

AutoIgnite

On, Off

Turns automatic ignition on or off.

AutoReIgnite

On, Off

Turns automatic re-ignition on or off.

ControlMode

Dual, SingleLeft, SingleRight

(DFID only) Sets the control mode. When the control mode is set to “Dual,” both left and right columns are connected. When it is set to “SingleLeft” or “SingleRight,”  either a left or right column is connected.

Extinguish

 

Extinguishes the flame.

Flame

On, Off

Reports whether  the flame is on or off (read-only).

H2Ctrl

On, Off

Turns the H2 gas flow control on or off.

H2Flow

0.0 to 1200.0 ml/min

Sets the H2 gas flow.

Ignite

 

Ignites the flame.

Polarity

Positive, Negative

(DFID only) Sets the detector polarity.

Special commands and properties for the Electron Capture Detector (ECD):

 

Command/Property

Range/Values

Description

Current

0.0 to 2.0 nA

Sets the detector current.

 

Special commands and properties for the Thermal Conductivity Detector (TCD) and Dual Thermal Conductivity Detector (DTCD):

 

Command/Property

Range/Values

Description

Current

0.0 to 200.0 mA

Sets the detector current.

Polarity

Positive, Negative

Sets the detector polarity.

PreHeaterTemp

 

(DTCD only) Controls the preheater temperature.

*.UpperLimit

0.00 to 400.00 °C

Sets the maximum preheater temperature.

*.Nominal

0.00 to 400.00 °C

Sets the nominal preheater temperature.

*.Value

0.00 to 400.00 °C

Indicates the actual preheater temperature (read-only).

PreAmp

On, Off

(DTCD only) Turns the 10x amplifier at the detector output ON or OFF.