Commands and Tips for Third-Party Devices
Perkin Elmer Series 200 UV/VIS Detector

Chromeleon provides the commands and parameters from the table below (please note that the display Filter level determines which commands and properties are displayed) to control the Series 200 UV/VIS Detector. Also, refer to the information in the detector's User's Manual.

General Commands and Properties

Command/Property

Description

Connect

Connects the device to Chromeleon

Connected

Indicates whether the instrument is connected to Chromeleon, that is, under computer control (also, see Connect, Disconnect)

Disconnect

Disconnects the instrument from Chromeleon and activates manual control from the instrument panel.

DriverID

Unambiguous driver name (read-only).

DriverVersion

Driver version (read-only).

FirmwareVersion

Indicates the firmware version of the instrument (read-only).

ModelNo

Reports the instrument's model number (read-only).

Manufacturer

Manufacturer of the driver (read-only).

SerialNo.

Indicates the serial number of the instrument (read-only).

 

 

Detector Commands:

Command/Property

Range/Value

Description

Autozero

 

Sets the value of the output signal to 0.

DisplayOperationTime

0-99999

Reports the number of hours the display has been on (read-only).

Lamp

On, Off

Switches the lamp on or off.

LampOperationTime

0-9999

Reports the number of hours the lamp has been on (read-only).

OperationTimeSinceCalibration

0-999

Reports the number of operating hours since the last calibration (read-only)

Ready

Ready, NotReady

Indicates whether the instrument is ready for injection (read-only).

 

UV_VIS Channel Commands:

Command/Property

Range/Value

Description

AcqOn/AcqOff

 

Start and end times for data acquisition.

DataCollectionRate

0.2 - 50.0 samples/s

Specifies the data acquisition rate in samples/seconds.

Delta

-3.11-3.11 mAU/s

Reports the signal's slope, i.e., the difference between the current value and the value one second ago (read-only). This is useful for triggers.

LampType

Deuterium (UV), Tungsten (VIS), Unknown

Reports the type of lamp installed in the detector (read-only).

PeakWidth

0.10 - 25.00s

Specifies the amount of smoothing to apply to the chromatogram. Depending on the setting, the RiseTime (see below) is changed automatically. Select a value:

0.10s (RiseTime = 0.02s)

0.25s (RiseTime = 0.05s)

0.50s (RiseTime = 0.10s)

1.00s (RiseTime = 0.20s)

2.50s (RiseTime = 0.50s)

5.00s (RiseTime = 1.00s)

10.0s (RiseTime = 2.00s)

25.0s (RiseTime = 5.00s)

Retention

 

Reports the retention time of the signal (read-only).

RiseTime

0.02 - 5.00s

Specifies how quickly the detector responds to a change in signal (read-only). The Rise Time defines how long it takes for an outgoing signal to rise from 10% to 90% of the end value.

A large rise time will reduce the short term noise, but will affect the peak height and symmetry.

Signal

 

Reports the current signal value [mAU] (read-only).

Wavelength

 

Sets the wavelength (the available range is from 190 to 700 nm).