Commands and Tips for Third-Party Devices
Perkin Elmer Autosystem (XL) and Clarus GCs: Inlet
The following commands and properties are available for inlet control (please note that the display Filter level determines which commands and properties are displayed):
Property |
Range/Values |
Description |
FollowOven |
On or Off |
(Only for PPS and POC injectors) If the property is set to On, the inlet temperature follows the oven temperature (+5°C). |
InletType |
PKD, CAP, GSV, PSSO, PSSI, POCI, POCO |
Indicates the inlet type (read-only): |
Ready |
Ready/NotReady |
Set to Ready when the inlet heating component is ready. (If a PPC is installed, the SplitStatus setting must also be Ready.) |
Status |
None |
Indicates the current status of the associated heating component. |
TempCtrl |
On or Off |
Enables or disables temperature control for the inlet. |
Temperature |
Without PTR: With PTR: |
Properties related to the column oven. Note: If a temperature is set, TempCtrl is automatically set to On. |
Temperature. |
Sets the nominal temperature. For the PSSI and POCI, this command sets a temperature ramp, instead. A maximum of two inlet temperature ramps are allowed. |
|
Temperature. |
Indicates the current temperature (read-only) |
In addition, Chromeleon supports the following commands and/or properties if the Autosystem GC includes a PPC:
Property |
Values |
Description |
PurgeCtrlMode |
Auto, Fixed |
Sets the purge flow compensation mode (level: Advanced or higher). (For more information, refer to the manual for the instrument.) |
PurgeFlow |
0.0 – 100.0 ml/min |
Sets the split flow offset value to compensate the reduction of the split flow due to septum purge flow. (The property can be set only if PurgeCtrlMode = Fixed.) |
SplitCtrlMode |
Flow, Ratio |
Sets the split control mode. Specifies whether the flow or the flow ratio is controlled. |
SplitFlow |
0 – 500 ml/min |
Sets the target value for the total split flow. (The property can be set only if SplitCtrlMode = Flow.) |
SplitRatio |
1-6000 |
Sets the target value for the split ratio. (The property can be set only if SplitCtrlMode = Ratio.) |
SplitStatus |
NotUsed, Ready, NotReady, Alarm, Unknown |
Indicates the current split status. |
TotalSplitFlow |
0 – 500 ml/min |
Indicates the current total flow (split flow + purge flow; read-only). |
For installation details, refer to: