Commands and Tips for Third-Party Devices
Perkin Elmer Autosystem (XL) and Clarus GCs: Inlet

The following commands and properties are available for inlet control (please note that the display Filter level determines which commands and properties are displayed):

 

Property

Range/Values

Description

FollowOven

On or Off

(Only for PPS and POC injectors) If the property is set to On, the inlet temperature follows the oven temperature (+5°C).

InletType

PKD, CAP, GSV, PSSO, PSSI, POCI, POCO

Indicates the inlet type (read-only):
PKD: Packed column
CAP: Capillary
GSV: Gas Sampling Valve
PSSO: Split/Splitless follow oven
PSSI: Split/Splitless
POCI: On Column
POCO: On Column follow oven

Ready

Ready/NotReady

Set to Ready when the inlet heating component is ready. (If a PPC is installed, the SplitStatus setting must also be Ready.)

Status

None
Ready
NotReady
FilamentOver
Unknown

Indicates the current status of the associated heating component.

TempCtrl

On or Off

Enables or disables temperature control for the inlet.

Temperature

Without PTR:
100 - 450°C or
50 - 450°C CAP/PKD
(50 - 350°C for GSV)

With PTR:
-99 - 500°C

Properties related to the column oven.

  Note:

If a temperature is set, TempCtrl is automatically set to On.

Temperature.
Nominal

Sets the nominal temperature. For the PSSI and POCI, this command sets a temperature ramp, instead. A maximum of two inlet temperature ramps are allowed.

Temperature.
Value

Indicates the current temperature (read-only)

 

In addition, Chromeleon supports the following commands and/or properties if the Autosystem GC includes a PPC:

 

Property

Values

Description

PurgeCtrlMode

Auto, Fixed

Sets the purge flow compensation mode (level: Advanced or higher). (For more information, refer to the manual for the instrument.)

PurgeFlow

0.0 – 100.0 ml/min

Sets the split flow offset value to compensate the reduction of the split flow due to septum purge flow. (The property can be set only if PurgeCtrlMode = Fixed.)

SplitCtrlMode

Flow, Ratio

Sets the split control mode. Specifies whether the flow or the flow ratio is controlled.

SplitFlow

0 – 500 ml/min

Sets the target value for the total split flow. (The property can be set only if SplitCtrlMode = Flow.)

SplitRatio

1-6000

Sets the target value for the split ratio. (The property can be set only if SplitCtrlMode = Ratio.)

SplitStatus

NotUsed, Ready, NotReady, Alarm, Unknown

Indicates the current split status.

TotalSplitFlow

0 – 500 ml/min

Indicates the current total flow (split flow + purge flow; read-only).

 

For installation details, refer to:

 Perkin Elmer Autosystem (XL) and Clarus GCs: Installation