Commands and Tips for Third-Party Devices
Gilson 202 Fraction Collector (Ext.)/204 Fraction Collector
The following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
Property |
(Min/Max) Values |
Purpose/Remarks |
ModelNo |
N/A |
Indicates the model number (for the Gilson 202 including the firmware version; (read-only). |
FirmwareVersion |
N/A |
Indicates the firmware version |
ValvePosition |
Possible values: |
Switches the fractionation valve to the Drain or Collect position (The property is available only if a fractionation valve is installed.) |
RackType |
Possible values: All Gilson 202 rack types |
Indicates the used rack type, specified in the Server Configuration program (read-only). |
TubePosition |
Range: 1 … n. |
Moves the collect head to the specified position. |
MovementMode |
Possible values: |
In Secure mode, the fractionation valve switches to the Drain position before the collect head starts moving and switches back to the Collect position when the tube position has been reached. If the fractionation valve is already in the Drain position, the instrument does not automatically switch to the Collect position. (The property is available only if a fractionation valve is installed.) |
Command |
Purpose/Remark |
Home |
Moves the collect head to the Home position (drain outlet) |
Next |
Moves the collect head to the next tube. If the collect head is positioned at the last tube, it moves to the first one. |
Drain |
If the rack type is code 10, the collect head is positioned over the central rack hole. For all other rack types, the fractionation valve is switched to the Drain position. (The property is available only if a fractionation valve is installed or if Code 10 is selected as the rack type.) |
Further Information
For information about how to install the Gilson 202 fraction collector with the Gilson 202 Fraction Collector (ext.) device driver and the Gilson 204 fraction collector with the Gilson 204 Fraction Collector device driver, refer to nstallation Instructions Gilson 202 Fraction Collector (Ext.)/204 Fraction Collector: Installation in the Administrator Help section.