Commands and Tips for Thermo Scientific Devices
Thermo Finnigan aQa Mass Spectrometer
TIC and SIM Channel Properties and Commands
These channels are standard data channels with the following properties (please note that the display Filter level determines which commands and properties are displayed):
Property |
Min |
Max |
Default |
Description |
MinMass |
2.00 |
1636.00 |
Read-only |
Minimum mass recorded in this trace. (SIM channels only) |
MaxMass |
2.00 |
1636.00 |
Read-only |
Maximum mass recorded in this trace. (SIM channels only) |
Polarity |
Negative |
Positive |
Read-only |
Polarity |
SourceVoltage |
0 V |
200 V |
Read-only |
aQa max voltage. (SIM channels only) |
SourceVoltage |
0 V |
200 V |
Read-only |
aQa max voltage range. (TIC/TICF channels only) |
BasePeakMode |
No |
Yes |
MS only |
Extracts Base Peak chromatograms during data acquisition |
Filter |
n/a |
n/a |
Read-only |
Filter string as used in Thermo Finnigan aQa method. |
Tip:
Support for data compression is not available. Thus, the Step, MaxAutoStop, and Average commands cannot be set. Manual data acquisition is not possible, either. In addition, acquisition is started at the time 0.000 and stopped at the end of the program for all channels that deliver data according to an aQa method file. These channels do not appear in the Program Wizard or on the control panel.
Device Properties and Commands
Property |
Min |
Max |
Default |
Description |
Ready |
Ready, Busy |
|
Read-only |
Indicates whether the spectrometer is busy or idle. |
Operation |
On, Off, Standby |
|
Off |
Indicates whether the spectrometer is busy or idle. |
Status |
n/a |
|
Read-only |
Indicates the spectrometer's operational status. |
Range |
1 |
10 |
|
Scaling range used for online signal plot during data acquisition. (Can be set only before data acquisition starts.) |
Smoothing |
None, Gaussian, Boxcar |
None |
|
Smoothing algorithm. Note: This parameter is used by Xcalibur raw data extraction during online and post-run channel extraction. |
SmoothingPoints |
3, 5, 7, 9, |
3 |
|
Number of data points used for smoothing. Note: This parameter is used by Xcalibur raw data extraction during online and post-run channel extraction. |
HardwareVersion |
n/a |
|
Read-only |
The mass spectrometer's hardware version. Note: This value is available only after the first raw data acquisition. |
FirmwareVersion |
n/a |
|
Read-only |
The spectrometer's firmware version. Note: This value is available only after the first raw data acquisition. |
Command |
Description |
Connects the device |
|
Disconnect |
Disconnects the device |
Resets the device to the default values. |