Commands and Tips for Third-Party Devices
CTC Analytics PAL Samplers: Tips for Headspace Operation

Chromeleon supports two headspace injection modes:

HS_Single:  Standard headspace injection

HS_Progr:  Progressive headspace injection

In HS_Single mode, the sampler performs a standard headspace injection, including sample preparation. You must specify an incubation time (IncubationTime parameter) for every sample in the sequence.

In HS_Progr mode, you do not have to explicitly set the incubation time for samples. Instead, you can increase the incubation time of the previous sample by a certain value (IncubOffset). The IncubTimeStart parameter determines the start of the incubation time of the first sample in a sequence. Thus, the incubation time for the sample number n is calculated as follows:

IncubTimeStart + (n-1) * IncubOffset .

 Tip:  

You are not allowed to change the IncubOffset or IncubTimeStart parameters in a sequence, i.e., they can be set before the start of the sequence or they must have the same value in each sample program that is used in the sequence.

In both modes, Chromeleon needs information about the minimum time difference between two sample injections. Use the GCCycleTime parameter to specify the time required by the GC after injection until the instrument is ready for the next injection.

 Tip:  

Make sure the cycle time is long enough for completion of all actions by the GC (injection procedure, chromatographic separation, cleaning, warming up of the inlet, etc.) and the sampler (cleaning before and after injection, preparation of new samples, etc.).

The PAL sampler automatically places the vials in the agitator for sample preparation. After the injection procedure, the vials are moved back to their original positions in the tray.

 Caution:  

Always run an emergency program to cool down the agitator (or other temperature-controlled components) before removing the samples manually from the agitator.

 Caution:  

Always run an emergency program to cool down the agitator (or other temperature-controlled components) before removing the samples manually from the agitator.

 Tip:  

The GC may require more time, e.g., to heat up the inlet, column, and/or detector until it has reached the new temperature. In this case, certain commands, such as Wait and Inject, delay the execution of the program, possibly even beyond the sample incubation time. Therefore, Dionex recommends setting the real time of the GC cycle to GCCycleTime and adding 30 to 60 seconds for safety reasons. (You might have to add more time, depending on the GC running conditions; for more information, refer to the GC instruction manual.)

 Note:  

The Audit Trail messages for sample preparation are displayed both in the currently running sample and in the sample that is prepared.

For more information, refer to  PGM File for HS_Progr Inject Mode.