ER Controller
In addition to the General Device Commands, the following commands are available for controlling the ERC 10 operating functions in an RFIC-ER system (please note that the display Filter level determines which commands are displayed).
Command/Property |
Description |
ER1 |
Note: The ER1 is the eluent purification column. |
*.StartDate |
The day on which the column was first used (read-only). |
*.ExpirationDate |
The last day on which the column can be used (read-only). |
*.PercentageUsed |
An indication of how long the column has been in operation, expressed as a percentage of the expected column lifetime (read-only). |
*.PercentageRemaining |
Estimated remaining lifetime before the column expires, expressed as a percentage of the expected column lifetime (read-only). |
ResetER1 |
Resets ER1.StartDate to the current date, ER1.ExpirationDate to StartDate + 365 days, and all other ER1 commands or properties to the initialization values. |
ER2 |
Note: The ER2 is the analyte trap column. |
*.StartDate |
The day on which the column was first used (read-only). |
*.ExpirationDate |
The last day on which the column can be used (read-only). |
*.InjectionsRemaining |
Estimated number of injections that can be done before the column expires (read-only). |
*.InjectionsDone |
Number of injections done since the column was first used (read-only). |
*.VolumeRemaining |
Estimated sample volume that can be injected before the column expires (read-only). |
*.VolumeUsed |
Total sample volume that has been injected since the column was first used (read-only). |
ResetER2 |
Resets ER2.StartDate to the current date, ER2.ExpirationDate to StartDate + 365 days, and all other ER2 commands or properties to the initialization values. |
EC |
Note: EC is the eluent counter. |
*.HoursUsed |
Total number of hours the eluent has been in use (read-only). |
*.HoursRemaining |
Estimated number of hours before the eluent expires (read-only). |
*.VolumeUsed
|
Total sample volume injected since the last eluent change (read-only). |
*.VolumeRemaining |
Estimated sample volume that can be injected before the eluent expires (read-only). |
*.InjectionsDone |
Total number of injections since the last eluent change (read-only). |
*.InjectionsRemaining |
Estimated number of samples that can be injected before the eluent expires (read-only). |
ResetEC |
Resets the eluent counter to zero. |
SetEcSuppressorStartup |
Sets the number of hours before the eluent expires to 16 hours (required during initial installation of the RFIC-ER system). |
TotalGasVolume |
The volume of gas detected within the past hour (read-only). |
ResetTotalGasVolume |
Resets the gas volume to zero. |