Ensuring System Wellness (IC Devices)
Conductivity Verification (ICS-1000/1500/2000 or ICS-1100/1600/2100)

Use this procedure to test the conductivity variance on the DS6 conductivity cell electronics.

 Note:

Do not open a System Wellness Control Panel if the name includes "Service.pan." These Wellness panels are reserved for use by Dionex Service Representatives.

  1. Open the System Wellness Control Panel for the ICS-1000/1500/2000 or ICS-1100/1600/2100 (see  Opening a Wellness Control Panel).

  2. Locate the Dummy Cell Test controls. Click Instructions for an overview of the test procedure.

  3. Click On to begin the test.

  4. When the test is complete, click Log to record the new dummy cell value in the Audit Trail. The new dummy cell value should be 21 mS; if it is not, the CPU card may be defective. Contact Dionex for assistance.